AneuGuidederived metal coverage ratio and pore density in Pipeline embolization devices: validation against VasoCT and association with aneurysm occlusion.
Yakun Chen, Rong Zou, Yejie Shan, Hao Gao, Bin Ren, Shengli Guo, Peng Wang, Jianning Zhang,
Senior Department of Neurosurgery, Chinese PLA General Hospital, Beijing, China. · ArteryFlow Technology Co., Ltd., Hangzhou, China.
Key points
From the paper's conclusion- AneuGuide provided MCR and PD estimates that agreed well with VasoCT-derived measurements. Higher MCR and PD were associated with aneurysm occlusion at short-term follow-up. These findings support the utility of AneuGuide for geometric assessment of FD deployment and suggest a potential role in treatment-related evaluation.
Abstract
BACKGROUND: Metal coverage ratio (MCR) and pore density (PD) are key factors affecting the outcomes of intracranial aneurysms (IAs) treated with flow diverters (FDs), yet their in vivo quantitative assessment remains challenging. This study evaluated the accuracy of AneuGuide-derived MCR and PD against VasoCT-based measurements and explored their association with short-term aneurysm occlusion. METHODS: This retrospective study enrolled patients treated with a single pipeline embolization device (PED) between June 2022 and April 2024. PED wires were reconstructed from postoperative VasoCT images to calculate MCRCT and PDCT at the aneurysm neck. Corresponding values (MCRAG and PDAG) were obtained using AneuGuide simulations. Agreement and correlation between the two methods were assessed using Bland-Altman analysis and Pearson correlation. The association between AneuGuide-derived parameters and aneurysm occlusion was evaluated using receiver operating characteristic (ROC) curves. RESULTS: A total of 48 aneurysms were included. AneuGuide showed good agreement with VasoCT for both MCR (r=0.65, P<0.001) and PD (r=0.79, P<0.001), with no significant differences (P>0.05). A strong correlation was observed between MCR and PD (r=0.868, P<0.001). Non-occluded aneurysms exhibited lower MCR (22.05%±3.02% vs. 27.14%±7.65%, P=0.034) and PD (13.53±3.85 vs. 17.50±6.28 pores/mm2, P=0.058). ROC analysis yielded AUCs of 0.770 for MCR and 0.742 for PD. CONCLUSIONS: AneuGuide provided MCR and PD estimates that agreed well with VasoCT-derived measurements. Higher MCR and PD were associated with aneurysm occlusion at short-term follow-up. These findings support the utility of AneuGuide for geometric assessment of FD deployment and suggest a potential role in treatment-related evaluation.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Safety, efficacy and angiographic outcomes of the Lattice flow diverter for intracranial aneurysms: A systematic review and meta-analysis.The neuroradiology journal · 2026
- MRA-based aneurysm occlusion assessment in the early phase after flow diverter treatment: a multicenter retrospective study focusing on the impact of aneurysm diameterNeurosurgical Review · 2026
- Efficacy and safety of flow diverters for the treatment of intracranial aneurysms: a systematic review and meta-analysis.Frontiers in surgery · 2026
- Multicenter experience of pipeline embolization device used in small caliber vessels (< 2 mm) for intracranial aneurysm treatment and mid-term resultsFrontiers in Neurology · 2026
- Closer matching between flow diverter diameter and parent artery diameter is associated with improved aneurysm occlusion.Surgical neurology international · 2026
- Ankyras simulation for flow diversion of intracranial aneurysms: a standardized workflow integrated clinical validation study.Journal of neurointerventional surgery · 2026