Closer matching between flow diverter diameter and parent artery diameter is associated with improved aneurysm occlusion.
Yoshimasa Fukui, Takashi Fujii, Kunihiro Matsumoto, Shuta Maehara, Kousei Maruyama, Kousuke Takigawa, Masahiro Yasaka, Hiroshi Aikawa,
Department of Neurosurgery, Fukuoka Neurosurgical Hospital, Fukuoka, Japan. · Department of Radiology, Fukuoka Neurosurgical Hospital, Fukuoka, Japan.
Key points
From the paper's conclusion- Greater FD elongation post-deployment may be linked to lower aneurysm occlusion rates at 6 months. Choosing an FD size that matches the parent vessel diameter is crucial for optimal occlusion.
Abstract
BACKGROUND: Selecting the flow diverter (FD) size is key to treatment outcomes in unruptured intracranial aneurysms. However, the impact of differences between the nominal and actual deployed device sizes on aneurysm occlusion is not fully understood. This study retrospectively evaluated the relationship between FD size and aneurysm occlusion. METHODS: We retrospectively analyzed 54 patients with unruptured intracranial aneurysms who underwent FD placement at our institution from March 2023 to March 2025, where both length (Length_D) and diameter (Size_D) discrepancies were evaluated. Devices used were a Flow Redirection Endoluminal Device in 31 cases and a Pipeline Embolization Device in 23 cases. Length_D is the difference between the nominal device length and actual deployed length, while Size_D is the difference between the nominal device diameter and actual deployed diameter. The primary endpoint was aneurysm occlusion at 6 months, assessed using the O'Kelly-Marotta (OKM) scale. Adequate occlusion was defined as OKM grades C-D and incomplete occlusion as OKM grades A-B. Patients were divided into adequate and incomplete occlusion groups, and Length_D and Size_D were compared. RESULTS: Adequate occlusion was achieved in 68.5% of patients. The mean Length_D was 2.9 ± 3.9 mm in the adequate occlusion group and 6.7 ± 5.1 mm in the incomplete occlusion group, with a significantly greater value in the incomplete occlusion group. CONCLUSION: Greater FD elongation post-deployment may be linked to lower aneurysm occlusion rates at 6 months. Choosing an FD size that matches the parent vessel diameter is crucial for optimal occlusion.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Factors Associated with Aneurysm Occlusion after Flow Diverter Treatment.Journal of neuroendovascular therapy · 2026
- MRA-based aneurysm occlusion assessment in the early phase after flow diverter treatment: a multicenter retrospective study focusing on the impact of aneurysm diameterNeurosurgical Review · 2026
- Very short flow diverters for intracranial aneurysms: technical feasibility, angiographic outcomes, and clinical safety across diverse cerebrovascular locations.Neuroradiology · 2026
- Efficacy and safety of flow diverter therapy for unruptured intracranial aneurysms compared to conventional endovascular therapy: a multicentre, randomised, open-label trial in South Korea.EClinicalMedicine · 2026
- Longitudinal volume evolution of large and giant intracranial aneurysms following endovascular flow diversion.Brain & spine · 2026
- Efficacy and safety of flow diverters for the treatment of intracranial aneurysms: a systematic review and meta-analysis.Frontiers in surgery · 2026