Ankyras simulation for flow diversion of intracranial aneurysms: a standardized workflow integrated clinical validation study.
Alan R Rheaume, Christine Denice, Seyed Morsal Mosallami Aghili, Michael Jean, Carla Guerrero, Lukasz Chojnowski, Daniel Kim, Tasnim Tanim,
Department of Neurosurgery, Stony Brook Medicine, Stony Brook, New York, USA. · Stony Brook Medicine, Stony Brook, New York, USA.
Key points
From the paper's conclusion- Using a robust 3D DynaCTA centerline based validation, Ankyras demonstrated high accuracy for PED deployment length and landing zone prediction within a standardized clinical treatment workflow of pre-procedural planning, intraoperative guidance, and post-deployment assessment.
Abstract
BACKGROUND: Accurate flow diverter (FD) sizing is critical for successful intracranial aneurysm (IA) treatment. We evaluated the predictive performance of the Ankyras FD simulation platform within a standardized simulation based clinical workflow using three-dimensional (3D) DynaCTA centerline analysis as an advanced validation reference standard. METHODS: A retrospective analysis of a prospectively maintained institutional database of Ankyras guided IA treatments between April 2024 and October 2025 was performed. 3D rotational angiography data generated Ankyras simulations for treatment planning, device selection, and intraoperative guidance. Simulated length (SL) and proximal landing zone (PLZ) were compared with post-deployment length (DL) from DynaCTA centerline measurements (3D Slicer). Accuracy was assessed using correlation coefficients, mean absolute error (MAE), and Bland-Altman analysis. RESULTS: 54 consecutive patients (48 women, mean age 54.2 years) with 63 IAs underwent 57 pipeline embolization device (PED) treatments. All achieved procedural success without intraoperative complications; two required adjunctive stent placement for malapposition. Mean SL was 21.44±3.53 mm and DL was 21.49±3.52 mm (n=46), with a strong SL-DL correlation (r=0.91), minimal bias (-0.05 mm), and 1.24 mm MAE. PLZ demonstrated a strong correlation (r=0.94), low bias (0.14 mm), and minimal MAE (1.34 mm) between Ankyras and the deployed values (n=45). Bland-Altman analysis showed narrow agreement limits without evidence of proportional error. CONCLUSIONS: Using a robust 3D DynaCTA centerline based validation, Ankyras demonstrated high accuracy for PED deployment length and landing zone prediction within a standardized clinical treatment workflow of pre-procedural planning, intraoperative guidance, and post-deployment assessment.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Hemodynamic-Structural Multi-Fidelity Assessment of Endovascular Flow Diversion Strategies for Cerebral Aneurysm Treatment: An Integrated Computational and Experimental FrameworkbioRxiv · 2026
- Safety, efficacy and angiographic outcomes of the Lattice flow diverter for intracranial aneurysms: A systematic review and meta-analysis.The neuroradiology journal · 2026
- AneuGuidederived metal coverage ratio and pore density in Pipeline embolization devices: validation against VasoCT and association with aneurysm occlusion.Quantitative imaging in medicine and surgery · 2026
- From Device Modeling to Blood Flow Simulation: A Study of the Woven EndoBridge.International journal for numerical methods in biomedical engineering · 2026
- Closer matching between flow diverter diameter and parent artery diameter is associated with improved aneurysm occlusion.Surgical neurology international · 2026
- First clinical experience with the surpass elite flow diverter in the treatment of intracranial aneurysmsNeuroradiology · 2026