Design assessment and preliminary performance evaluation of a novel device for non-invasive emergency treatment of acute respiratory failure.
Federica Ciceri Negri, Eleonora Salurso, Giuseppe Giardina, Filippo Consolo, Giovanni Landoni, Alberto Cesare Luigi Redaelli, Gianfranco Beniamino Fiore
Department of Electronics, Information and Bioengineering, Politecnico Di Milano, Milan, Italy. · Department of Anaesthesia and Intensive Care, IRCCS San Raffaele Scientific Institute, Milan, Italy.
Key points
From the paper's conclusion- Our results confirmed the validity of our design solutions and suggest a potentially safe and efficient profile as a promising solution for early CPAP use in emergency settings.
Abstract
UNLABELLED: Acute respiratory failure (ARF) requires rapid intervention to prevent patient deterioration and life-threatening complications. Early respiratory support through prehospital Continuous Positive Airway Pressure (CPAP) has been shown to improve ARF outcomes. However, existing CPAP devices present technical barriers for use in out-of-hospital emergency scenarios. This study presents the design and preliminary performance evaluation of a novel portable, ready-to-use CPAP device for out-of-hospital respiratory support. The device includes a blower providing steady air insufflation through a face mask, with Positive End-Expiratory Pressure (PEEP) regulated by two fixed-point PEEP valves. To identify optimized operational conditions ensuring target respiratory support, a lumped-parameter model (LPM) was developed to simulate device-patient interaction. Input data for the LPM included experimental characteristics of the device components. Simulations were conducted using breathing profiles representing varying ARF-like conditions and severity. Additionally, computational fluid dynamics (CFD) analysis was performed to assess the risk of CO2 rebreathing through the mask. LPM simulations described the dynamic interaction between the breathing profile and device performances, and the consequent pressure and flow patterns during the respiratory cycle. Across all simulated conditions and device settings, PEEP remained close to the predefined target of 5 cm H2O without reaching critical thresholds, confirming design choices efficacy. CFD analysis indicated an end-expiratory residual CO2 concentration of ~2%, suggesting a low risk of rebreathing between cycles. Our results confirmed the validity of our design solutions and suggest a potentially safe and efficient profile as a promising solution for early CPAP use in emergency settings. SUPPLEMENTARY INFORMATION: The online version contains supplementary material available at https://doi.org/10.1007/s13534-026-00619-3.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Positive end-expiratory pressure versus sham valve/zero end-expiratory pressure in cardiopulmonary resuscitation during manual ventilation to improve neurological outcomes in adult patients suffering an out-of-hospital cardiac arrest - an investigator-initiated, pragmatic, registry-based, multicenter, parallel-group, triple-blind randomized controlled superiority clinical trial in the ARREST registry (REVIVE-PEEP protocol Stage-1 Registered Report)medRxiv · 2026
- In Vitro Comparison of a Low-Cost Bubble-NIV and Ventilator-Driven NIV.Respiratory care · 2026
- A clinical guide to non-invasive respiratory support in acute respiratory failure: ventilation settings, technical optimization and clinical indicationsCritical Care · 2025
- VENTILA2: A Fuzzy Logic-Based Simulation and Decision-Support Framework for Pressure-Controlled Ventilation-A Proof of Concept.Healthcare (Basel, Switzerland) · 2026
- Disposable vs. conventional NPPV in acute respiratory acidosis.Open medicine (Warsaw, Poland) · 2026
- Simple and easy to build mechanical ventilator for pandemic health crises in resource-limited areas.Frontiers in public health · 2026