Prehospital Pediatric Ventilation Quality Following Use of a Feedback Device: A Randomized Study.
Joseph D Finney, Jeffrey Siegler, Jinli Wang, Kavya John, Debra Robinson, Thomas Grawey, Brad McClain, Sang Hoon Lee,
Department of Pediatrics, Washington University School of Medicine in Saint Louis, St. Louis, Missouri. · Department of Emergency Medicine, Washington University School of Medicine in Saint Louis, St. Louis, Missouri.
Key points
From the paper's conclusion- In our simulation study, EMS clinicians rarely met manual ventilation guidelines for pediatric patients in respiratory arrest. The use of a VFD substantially improved guideline adherence, suggesting broad potential to improve ventilation performance.
Abstract
OBJECTIVE: Manual ventilation is a lifesaving intervention for children in respiratory arrest, yet the quality of pediatric ventilation delivered by emergency medical services (EMS) clinicians across geographically distinct sites is poorly characterized. We evaluated baseline EMS ventilation performance and whether a ventilation feedback device (VFD) improves adherence to national and international ventilation guidelines for ventilation rate, tidal volume, and mask seal. METHODS: We conducted a single-anonymized, randomized, crossover study of EMS clinicians from 5 EMS agencies across 3 metropolitan sites in the United States. Using a simulated 3-year-old toddler manikin, participants completed 2 sequential 2-minute respiratory arrest scenarios separated by a 10-minute washout period. Participants performed ventilation with and without real-time VFD feedback in randomized order. The EOLifeX VFD recorded ventilation rate, tidal volume, and mask leak with each breath. Adherence to age-based guidelines was analyzed using generalized linear mixed-effects models with participants nested within sites. RESULTS: We enrolled 123 participants across 5 agencies. With VFD feedback, guideline-recommended tidal volumes were achieved in 87% of breaths compared with 16.2% without feedback. Ventilation rate adherence improved from 65% to 88.6%. Mask leak of less than 30% occurred in 88.4% of ventilations with feedback vs 80% without feedback. CONCLUSION: In our simulation study, EMS clinicians rarely met manual ventilation guidelines for pediatric patients in respiratory arrest. The use of a VFD substantially improved guideline adherence, suggesting broad potential to improve ventilation performance.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Accuracy of ventilation feedback devices compared with a reference standard during asynchronous ventilation with continuous chest compressions in a simulated cardiac arrest: a prospective non-randomized cross-over study.Resuscitation plus · 2026
- Real-time feedback during bag-valve-mask ventilation for out-of-hospital cardiac arrest: a before-and-after study (VEntilation during Cardiac Arrest Resuscitation Study, VECARS-2).Resuscitation · 2026
- Real-world adoption and use of a ventilation feedback device during out-of-hospital cardiac arrest: a population-based observational study from a French prehospital system.Resuscitation plus · 2026
- Impact of real-time respiratory function monitoring on neonatal mask ventilation training: a multicentre simulation-based crossover study.BMJ paediatrics open · 2026
- Agreement and Reliability of a Newly Developed Tidal Volume Monitoring Device in Bag-Valve Ventilation and Intubation Scenarios: A Simulation-Based Study.Bioengineering (Basel, Switzerland) · 2026
- Residents' Ventilator Management Performance in an Online Pediatric Acute Respiratory Distress Syndrome Case Scenario: Secondary Analysis of the Game-Based Education in Residency Study, 2015-2017.Pediatric critical care medicine : a journal of the Society of Critical Care Medicine and the World Federation of Pediatric Intensive and Critical Care Societies · 2026