Journal of Laboratory Medicine

Implementation and clinical concordance of a plasma pTau217 RUO assay within an in-house IVD/LDT workflow

Alexander Lenard, Lisa Bruch, Steffen Busch, Peter Findeisen

Key points

From the paper's conclusion
  • Conclusions Plasma pTau217 showed strong concordance with CSF Aβ42/40 status and potential for three-zone triage in memory-clinic patients.

Abstract

Objectives To describe implementation of an automated plasma phosphorylated tau 217 (pTau217) reagent in a laboratory-defined workflow and evaluate concordance with cerebrospinal fluid (CSF) amyloid status. Methods We analyzed 73 matched CSF-plasma pairs from a memory-clinic cohort using CSF Aβ42/40 as reference. Published thresholds defined three classification zones. Renal analyses within this plasma-only cohort were exploratory and included amyloid-status adjustment. Analytical observations comprised one plasma dilution series, single-sample stability testing, and serum/CSF precision data. Results Plasma pTau217 discriminated CSF Aβ-positive status with an area under the curve (AUC) of 0.918 (95 % CI 0.825–0.987). Exploratory AUC estimates were 0.957 (95 % CI 0.889–0.999) for estimated glomerular filtration rate (eGFR) ≥60 and 0.692 (95 % CI 0.250–1.000) for eGFR <60 mL/min/1.73 m 2 . The latter subgroup comprised 13 Aβ-positive and four Aβ-negative patients; no formal AUC comparison was performed. At the rule-out threshold, 25/73 patients were classified as negative, including two Aβ-positive cases (negative predictive value 0.920). The rule-in zone comprised 39/73 patients, including three Aβ-negative cases (positive predictive value 0.923); nine patients (12.3 %) were intermediate. Conclusions Plasma pTau217 showed strong concordance with CSF Aβ42/40 status and potential for three-zone triage in memory-clinic patients. Exploratory renal findings warrant confirmation in larger cohorts.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers

Implementation and clinical concordance of a plasma pTau217 RUO assay within an in-house IVD/LDT workflow — Science Skim