Clinical Chemistry and Laboratory Medicine (CCLM)Intraclass correlation

Analytical agreement and platform-specific decision thresholds for plasma p-tau217 measured on Lumipulse G600II and Cobas e801 in a paired CSF–plasma cohort

María Martínez-Bujidos, Alex Menéndez, Jose Arnau Pulido-Gracia, Dolores Vilas, Daniela Samaniego, Lourdes Ispierto, Cristian Morales-Indiano, Pau Pástor

Sociedad Española de Medicina de Familia y Comunitaria · Hospital Universitari Germans Trias i Pujol

Observational · n=157

Key points

From the paper's conclusion
  • Plasma p-tau217 demonstrates high analytical concordance and comparable diagnostic performance across automated platforms despite systematic concentration differences.

Abstract

OBJECTIVES: Plasma phosphorylated tau at threonine 217 (p-tau217) has emerged as a highly accurate blood-based biomarker of Alzheimer's disease (AD) pathology. As disease-modifying anti-amyloid therapies enter clinical practice, scalable biomarkers with robust and clinically interpretable decision thresholds are required. However, evidence on inter-platform comparability and threshold transferability across automated assays remains limited. METHODS: e801 - for plasma p-tau217 measurement in 157 consecutive patients undergoing lumbar puncture. Amyloid status was defined by the CSF Aβ42/Aβ40 ratio. Agreement was assessed using intraclass correlation coefficients and Bland-Altman analysis. Diagnostic performance was evaluated using receiver operating characteristic curves. Optimal thresholds were derived using the Youden index. Predefined rule-out (≥95 % sensitivity) and rule-in (≥95 % specificity) thresholds were explored, alongside alternative ≥90 % thresholds. RESULTS: Agreement between platforms was excellent (Spearman ρ=0.922; ICC(3,1)=0.922), although Bland-Altman analysis revealed a small systematic difference in absolute concentrations. Both assays showed comparable diagnostic accuracy for amyloid positivity (AUC=0.923 for both platforms; DeLong p>0.99), but required platform-specific thresholds. Rule-out and rule-in thresholds achieved ≥95 % sensitivity and specificity, with strong likelihood ratios and excellent categorical agreement (weighted κ=0.870). Approximately 30 % of individuals were classified in the grey zone. Using ≥90 % thresholds reduced the grey zone to 9-13 % while maintaining excellent agreement. CONCLUSIONS: Plasma p-tau217 demonstrates high analytical concordance and comparable diagnostic performance across automated platforms despite systematic concentration differences. Platform-specific dual-threshold strategies may support structured and clinically interpretable implementation, pending prospective multicenter validation.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers