Development of a measurement device and a lower leg phantom for standardized biomechanical testing of ankle braces
Jörg Schroeter, Robert Wendlandt, Arndt-Peter Schulz
Key points
From the paper's conclusion- The results demonstrate that the combination of the precise measurement device and the twoaxis phantom accurately simulates the biomechanical behavior of the human ankle joint, thereby enabling standardized, subject-independent orthosis testing.
Abstract
Orthoses are frequently used for the prevention and rehabilitation of ankle injuries. A diverse variety of products exists, but objective information comparing their mechanical support is often lacking. To objectively and reproducibly quantify the mechanical supportive effect of ankle braces, a highly precise measurement device for recording passive joint stiffness was developed alongside a lifelike two-axis silicone phantom. Because biomechanical in vivo tests are subject to strong inter-individual fluctuations, the use of measurement phantoms is indispensable for isolating purely mechanical stabilization properties. The measurement device was validated using a standardized mechanical spring phantom, while the two-axis silicone phantom was calibrated and validated against native in vivo stiffness values derived from a study of healthy subjects. The results demonstrate that the combination of the precise measurement device and the twoaxis phantom accurately simulates the biomechanical behavior of the human ankle joint, thereby enabling standardized, subject-independent orthosis testing.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Biomechanical evaluation of the stabilizing effect of wrist orthoses and cast treatments on an arm phantomCurrent Directions in Biomedical Engineering · 2026
- Wireless Instrumented Ankle Foot Orthosis (AFO) for Gait Cycle MonitoringInstruments · 2026
- The variable stiffness orthosis: customizable mechanics for assistance and rehabilitationJournal of NeuroEngineering and Rehabilitation · 2026
- A Pneumatic McKibben Muscle-Based Limb Volume Phantom Model for Wearable Strain Sensor Validation.Biomedical physics & engineering express · 2026
- Methods for the identification of ankle joint dynamics using a novel sub-malleolar interface.Clinical biomechanics (Bristol, Avon) · 2026
- Contact Mapping Using a Low-Cost Velostat Sensor Array in a Synthetic Ankle Model.Sensors (Basel, Switzerland) · 2026