From metrological comparability to clinical continuity: authorising laboratory result reporting when changing measurement procedures.
Raúl Rigo-Bonnin, David Ceacero-Marín, Míriam Valbuena-Asensio, Virgínia Mas-Bosch, Aurora Blanco-Font
Laboratori Clínic, IDIBELL, Hospital Universitari de Bellvitge, L'Hospitalet de Llobregat, Barcelona, Spain.
Key points
From the paper's conclusion- Changes in measurement procedure should culminate in an authorised reporting rule integrating metrological, analytical, clinical and risk evidence. Regression alone cannot establish continuity or interchangeability.
Abstract
OBJECTIVES: To develop and apply an operational framework integrating metrological, analytical, clinical and risk evidence to authorise an output-specific reporting rule after a change in measurement procedure. METHODS: Four measurement procedure changes were evaluated: alanine aminotransferase (ALT) procedures with and without pyridoxal-5'-phosphate activation; follicle-stimulating hormone (FSH) reagent replacement; routine vs. short-turnaround-time (STAT) parathyroid hormone (PTH) procedures; and tacrolimus procedure replacement with retention of the previous procedure for contingency use. The pathway integrated metrological comparability and compatibility, operational equivalence, controlled correction to a target reporting scale, independent clinical verification, uncertainty-aware classification, risk assessment and surveillance. RESULTS: The framework produced four qualitatively distinct outcomes: no cross-procedure continuity for ALT; direct continuity for raw FSH results; PTH continuity only through a corrected STAT result; and routine use of the new tacrolimus procedure, with the previous one restricted to contingency testing. FSH showed 100 % pair-level compatibility (95 % CI 91.2-100 %) and nominal percentage of clinically acceptable concordance (PCAC) of 96.6 % (82.2-99.9 %). PTH raw compatibility was assumption-sensitive at 90.0 % (76.3-97.2 %) vs. 87.5 % (73.2-95.8 %); the corrected output increased compatibility to 95.0 % (83.1-99.4 %), with nominal PCAC of 100 % (83.2-100 %). Tacrolimus met the mean-difference criterion but showed excess residual scatter, raw compatibility of 70.0 % (53.5-83.4 %) and nominal PCAC of 65.0 % (40.8-84.6 %); exploratory correction was not authorised. CONCLUSIONS: Changes in measurement procedure should culminate in an authorised reporting rule integrating metrological, analytical, clinical and risk evidence. Regression alone cannot establish continuity or interchangeability.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Verification of Patient Result Comparability Among Nine Hematology Analyzers Using the CLSI EP31-A-IR Guideline: A Four-Year Experience.International journal of laboratory hematology · 2026
- Qualitative results in proficiency testing: a systematic review of evaluation practices and error integration.Clinical chemistry and laboratory medicine · 2026
- The Laboratory Medicine Inference Framework: evidentiary boundaries from measurement to clinical action.Clinical chemistry and laboratory medicine · 2026
- When patient change and interpretive applicability diverge: a two-axis framework for laboratory medicine.Clinical chemistry and laboratory medicine · 2026
- Numerical and Interpretive Comparability of Optical and Mechanical Coagulation Analyzers: A Zlog-Supported Method Comparison Study.Diagnostics (Basel, Switzerland) · 2026