The Dutch colon cancer screening programme as an example of adopting indirect-outcome-based analytical performance specifications using a virtual analyser concept.
Marith Van Schrojenstein Lantman, Anja De Jong, Petra Van Der Groep, Evelien A Kemper, Monique De Groot, Renée Vossen, Karin Mohrmann, Joop Ten Kate,
SKML, Nijmegen, The Netherlands. · Bevolkingsonderzoek Nederland, Utrecht, The Netherlands.
Key points
From the paper's conclusion- Here, APS were based on a clinically consequential intermediate outcome, i.e. analytically induced variability in colonoscopy referral rate (±0.5 %), and implemented as a virtual analyser system, dividing the APS budget of 10 % equally across four sources of FIT variation: reagent/calibrator lot variation, sampling buffer lot variation, within-analyser variation, and between-analyser variation.
Abstract
To ensure reported laboratory results are of adequate quality, it is paramount that the set analytical performance specifications (APS) are tailored to the intended uses of the measurand. For measurands with a pivotal position in the clinical pathway, APS should be set based on clinical outcome. This paper, written as a result of an outcome-based-APS-summit in Prague, gives practical guidance on how to set outcome-based APS. Two approaches are mentioned with distinct starting points of the required performance (how much clinical misclassification do we allow, and which APS fulfils that premise) or the current performance (what is the effect of the current analytical performance characteristics (APC) on how much misclassification occurs). Importantly, APS should not just be set, but also requires monitoring, which method depends on the utilization of the test in clinical practice. The laboratory can monitor APS for so-called mother-daughter systems, individual analysers or a virtual-analyser system. This paper discusses the pros and cons of these approaches. To highlight the scope of clinically consequential outcomes for which APS can be set and monitored, the Faecal Immunochemical Testing for human haemoglobin (FIT), used for the Dutch Colorectal Cancer Screening programme (CRC-Screening), is used as an example. Here, APS were based on a clinically consequential intermediate outcome, i.e. analytically induced variability in colonoscopy referral rate (±0.5 %), and implemented as a virtual analyser system, dividing the APS budget of 10 % equally across four sources of FIT variation: reagent/calibrator lot variation, sampling buffer lot variation, within-analyser variation, and between-analyser variation.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- The Prague 2025 consensus checklist for reporting indirect outcome studies of Model 1b analytical performance specifications.Clinical chemistry and laboratory medicine · 2026
- Analytical performance specifications in the clinical laboratory: Historical evolution, conceptual foundations and a practical decision framework.Clinical biochemistry · 2026
- Analytical Performance Specifications Derived from External Quality Assessment Data: Comparison with Biological Variation-Based Models for Intermediate Precision Assessment.Annals of laboratory medicine · 2026
- Model 1b analytical performance specifications - report on the Prague 2025 workshop.Clinical chemistry and laboratory medicine · 2026
- Integrating APCS Score With Dual Biomarker Testing (FIT + mSEPT9) for Precision Risk Stratification and Predictive Modeling in High-Risk Populations for Colorectal Cancer.Annals of gastroenterological surgery · 2026
- B-077 Multicenter Comparison of PSA Immunoassays: Analytical Bias and Its Impact on Prostate Cancer ScreeningClinical Chemistry · 2026