An LR-based risk stratification framework for assessing interlaboratory variability in HEp-2 IFA pattern interpretation.
Chaochao Zhang, Yingxin Dai, Dan Cao, Rong Chen, Zhiyuan Gao, Yifan Gong, Anni Guo, Xiaowei Huang,
Department of Laboratory Medicine, Renji Hospital, Shanghai Jiao Tong University School of Medicine, Shanghai, China. · Department of Laboratory Medicine, Shanghai Public Health Clinical Center, Shanghai, China.
Key points
From the paper's conclusion- HEp-2 IFA interpretation variability differs in both frequency and potential diagnostic relevance. The LR-based framework and PDS complement conventional agreement measures by identifying discrepancies that alter diagnostic risk classification, potentially supporting risk-oriented proficiency assessment and targeted quality improvement.
Abstract
OBJECTIVES: Interlaboratory variability in HEp-2 indirect immunofluorescence assay (IFA) interpretation remains a major challenge. Conventional agreement metrics quantify disagreement but do not distinguish discrepancies by diagnostic relevance. We aimed to develop a likelihood ratio (LR)-based framework for clinically informed assessment of interpretation variability. METHODS: Pattern-specific positive LRs were derived from 33,690 routine HEp-2 IFA records and used to classify patterns into four risk tiers. A pattern discrepancy score (PDS) was developed to quantify discrepancies according to risk-tier displacement. Interlaboratory variability was assessed using a standardized 200-specimen panel tested by 18 laboratories, and diagnostic performance was explored in six laboratories. RESULTS: Pattern disagreement rates ranged from 17.5 to 50.0 %. PDS strongly correlated with disagreement rate (Spearman's ρ=0.936, p<0.001), but laboratories with similar disagreement rates differed in cross-tier discrepancy distribution. Most discrepancies remained within the same risk tier; among discordant interpretations involving high-risk reference patterns, 20.4 % crossed into another tier. Pattern concordance varied markedly, while titer agreement was moderate to almost perfect (weighted κ, 0.569-0.821). Across six laboratories, diagnostic performance varied (Youden Index, 0.391-0.617), but was not significantly associated with interpretation consistency. CONCLUSIONS: HEp-2 IFA interpretation variability differs in both frequency and potential diagnostic relevance. The LR-based framework and PDS complement conventional agreement measures by identifying discrepancies that alter diagnostic risk classification, potentially supporting risk-oriented proficiency assessment and targeted quality improvement.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- B-140 Performance and Workflow Impact of Automated Image Analysis in High-Volume ANA HEp-2 IIF ScreeningClinical Chemistry · 2026
- Evaluating diagnostic accuracy and analytical performance of HELIOS in detecting autoimmune disorders using external proficiency data as a tool: A three-year analysis.Annals of clinical biochemistry · 2026
- Upstream of likelihood ratio: A diagnostic framework for quantifying interpretive variability in forensic DNA mixture interpretation.Forensic science international. Genetics · 2026
- B-179 Quantifying the Quality of HEp-2 Cell Slides to Ensure the Accuracy of Indirect Immunofluorescence Antinuclear Antibody Testing ResultsClinical Chemistry · 2026
- Deep learning for automated classification of antinuclear antibody patterns on HEp-2 indirect immunofluorescence images.Immunologic research · 2026
- B-075 Can DIFF Scattergrams Predict Acute Leukemia Phenotype? A Standardized Interpretation Approach Using Mindray BC-6800PlusClinical Chemistry · 2026