B-075 Can DIFF Scattergrams Predict Acute Leukemia Phenotype? A Standardized Interpretation Approach Using Mindray BC-6800Plus
R I C A R D O RODRIGUEZ, Nancy Herrera, Roberto Vilca
Key points
From the paper's conclusion- Conclusion Standardized DIFF scattergram interpretation using the Mindray BC-6800Plus showed high overall agreement (80.8%) with flow cytometry immunophenotyping for acute leukemia phenotype classification in newly suspected cases at INEN.
Abstract
Background Early recognition of acute leukemia phenotype (lymphoid vs. myeloid) is essential for rapid diagnostic workflow and timely initiation of confirmatory testing. Automated hematology analyzers such as Mindray BC-6800Plus generate DIFF scattergrams that may suggest lineage patterns based on cellular distribution and abnormal clusters characteristics. However, standardized interpretation approaches remain limited, and their diagnostic performance compared with flow cytometry immunophenotyping has not been fully established. This study evaluated the diagnostic concordance and predictive performance of DIFF scattergram interpretation to classify acute leukemia phenotypes in newly suspected cases. Methods A cross-sectional diagnostic accuracy study was conducted at the Hematology Laboratory of the Instituto Nacional de Enfermedades Neoplásicas (INEN), Peru. Newly suspected acute leukemia cases (without age restriction) were collected between January 1, 2026, and February 10, 2026. DIFF scattergrams obtained from the Mindray BC-6800Plus analyzer were interpreted using a standardized visual approach (a red line was drawn parallel to the y-axis, where the debris area ends, as shown in the attached figure) and classified as suggestive of either lymphoid or myeloid phenotype. Flow cytometry immunophenotyping was used as the reference standard. A 2×2 contingency table was constructed. Sensitivity, specificity, positive predictive value (PPV), negative predictive value (NPV), and overall agreement were calculated. Results A total of 26 new suspected acute leukemia cases were analyzed. By immunophenotyping, 18 cases were lymphoid, and 8 cases were myeloid. DIFF scattergram interpretation classified:13 cases as lymphoid13 cases as myeloid Agreement between DIFF scattergram and immunophenotype, overall concordance was 80.8% (21/26). Performance for detecting Myeloid phenotypeSensitivity: 100% (8/8)Specificity: 72.2% (13/18)PPV: 61.5% (8/13)NPV: 100% (13/13)Accuracy: 80.8% (21/26) No myeloid cases were misclassified as lymphoid. However, 5 lymphoid cases (27.8%) were classified as myeloid by DIFF scattergram interpretation. Conclusion Standardized DIFF scattergram interpretation using the Mindray BC-6800Plus showed high overall agreement (80.8%) with flow cytometry immunophenotyping for acute leukemia phenotype classification in newly suspected cases at INEN. The method demonstrated excellent sensitivity and NPV (100%) for myeloid phenotype, suggesting it may be a useful screening tool to rapidly identify potential myeloid acute leukemia and prioritize confirmatory testing. However, a proportion of lymphoid cases were misclassified as myeloid, indicating that DIFF scattergrams should be used as an early supportive tool rather than a definitive lineage classifier.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Role of Gray Scattergram in Detecting Blasts Compared with Microscopic Peripheral Smear in Acute Leukemia Patients.Journal of blood medicine · 2026
- Utility of WBC Scatterplots and Suspect Flags Generated by Beckman Coulter LH-750 Hematology Analyzer in the Characterization of Leukemias and Related Hematological Malignancies.Indian journal of hematology & blood transfusion : an official journal of Indian Society of Hematology and Blood Transfusion · 2026
- Diagnostic agreement between ChatGPT-based interpretation of laboratory-generated flow cytometry graphical outputs and hematologist consensus in suspected acute leukemia.Cytometry. Part B, Clinical cytometry · 2026
- Beyond Analytical Agreement: Integrating Analytical Performance and Algorithm-Assisted Morphological Interpretation in a Comparative Evaluation of the Mindray BC-7800/BC-7900 and Sysmex XN-9100 Hematology Platforms.Diagnostics (Basel, Switzerland) · 2026
- Diagnostic Concordance and Superiority of Flow Cytometry Over Immunohistochemistry in Acute Leukemia Immunophenotyping: A Single-Center Study.Medical science monitor : international medical journal of experimental and clinical research · 2026
- Diagnostic accuracy of automated hematology analyzer abnormal flags for detecting hematological malignancies: A systematic review and meta-analysisPLoS ONE · 2026