Mass spectrometry-based screening for monoclonal gammopathy in individuals undergoing routine physical examinations: Analyses of consecutive 15,068 cases in China.
Zi-Han Yang, Li-Xin Wang, Yu-Xi Wang, Qian Di, Wei Su, Jian-Hua Han, Jian Li
Department of Hematology, Peking Union Medical College Hospital, Chinese Academy of Medical Sciences & Peking Union Medical College, Beijing, China. · Intelligene Biosystems Co., Ltd., Qingdao, Shandong, China.
Key points
From the paper's conclusion- The iMS-LC Assay demonstrated a higher detection yield than SPE for M-protein screening and identified additional low-abundance M-proteins, particularly in younger age groups.
Abstract
BACKGROUND: The intact M-protein Screening-Light Chain Assay (iMS-LC Assay) is a MALDI-TOF mass spectrometry (MS)-based method for M-protein detection without antibody enrichment. This study evaluated its feasibility for M-protein screening in comparison with conventional serum protein electrophoresis (SPE). METHODS: We prospectively enrolled 15,068 consecutive individuals aged ≥30 years undergoing routine physical examinations at Peking Union Medical College Hospital between December 2, 2024, and April 10, 2025. Residual serum samples from SPE were subjected to automated iMS-LC Assay screening. Samples positive by either MS or SPE underwent verification by serum immunofixation electrophoresis (IFE), free light chain (FLC) analysis, and MS isotyping. RESULTS: MS screening identified 201 positive cases compared to 70 by SPE (p < 0.001), demonstrating consistently higher detection rates across all age groups and more than a threefold increase among individuals under 60. MS positivity rates increased with age: males 30-39y (0.44%), 40-49y (1.19%), 50-59y (1.60%), 60-69y (2.51%), 70-79y (7.56%), ≥80y (14.29%); females 30-39y (0.44%), 40-49y (1.10%), 50-59y (1.47%), 60-69y (2.46%), 70-79y (3.92%), ≥80y (8.00%). CONCLUSION: The iMS-LC Assay demonstrated a higher detection yield than SPE for M-protein screening and identified additional low-abundance M-proteins, particularly in younger age groups.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Current Evidence for Integrating Mass Spectrometry into Clinical M-Protein Assessment in Plasma Cell Dyscrasias.Clinical chemistry · 2026
- One clone, one mass: The rise of mass spectrometry in the diagnosis and monitoring of monoclonal gammopathies.Clinica chimica acta; international journal of clinical chemistry · 2026
- Performance evaluation of an automated mass spectrometry system for the measurement and isotyping of monoclonal immunoglobulinsClinica Chimica Acta · 2026
- Harmonizing M-protein analysis: an international, multi-center reproducibility study of mass spectrometry and evaluation of concordance to electrophoretic methods.Clinical chemistry and laboratory medicine · 2026
- Automated peak detection, isotyping and quantitation of M-proteins by MALDI-TOF MS for replacement of immunofixation and serum protein electrophoresisJournal of Immunological Methods · 2026
- B-039 Harmonizing M-protein analysis: an international, multi-center reproducibility study of the EXENT® System and evaluation of concordance to electrophoretic methodsClinical Chemistry · 2026