A structured productivity challenge study of a fully automated mass spectrometry system
Arber Rexhaj, Katharina Habler, Bernd H. Northoff, Florian Arend, Michael Vogeser
Ludwig-Maximilians-Universität München
Key points
From the paper's conclusion- Under a high-throughput evaluation protocol in a routine clinical laboratory setting, the fully automated MS methods demonstrated performance and productivity characteristics consistent with the manufacturer's representative data, particularly with regard to calibration stability, precision, analytical throughput, and walk-away time.
Abstract
OBJECTIVES: i601 is a fully automated mass spectrometry analyzer designed for 24 h random-access operation in routine medical laboratories. We aimed to systematically assess its throughput, robustness, and precision - along with its ease of use and walk-away time - within a high-volume clinical laboratory setting using the first commercially available reagents. METHODS: reagents: Ionify steroids 1, Ionify steroids 2, Ionify vitamin D 25 and 24,25), was quantified using aliquots of three human serum pools. Identical series layouts were applied across four study days. After loading reagents, consumables, and samples and initiating the respective series, no further operator intervention was required until completion of the run. RESULTS: Across the four series, a total of 3,016 valid quantitative results were generated. Each series was completed within 6 h without operator intervention, corresponding to the minimum observed walk-away time. This equates to an average throughput of 119 analyses per hour. Analysis of the human serum pools demonstrated coefficients of variation (CVs) below 4 % (with few exceptions). CONCLUSIONS: Under a high-throughput evaluation protocol in a routine clinical laboratory setting, the fully automated MS methods demonstrated performance and productivity characteristics consistent with the manufacturer's representative data, particularly with regard to calibration stability, precision, analytical throughput, and walk-away time.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Comparison of six serum analyte workflows using routine liquid chromatography-tandem mass spectrometry methods at multiple laboratories with the Cobas i 601 analyzer, an automated mass spectrometry systemClinical Chemistry and Laboratory Medicine (CCLM) · 2026
- B-365 Breaking Throughput Barriers: Automated Vitamin D Testing on the Roche Cobas i601 LC-MS/MS in a Tertiary Hospital LaboratoryClinical Chemistry · 2026
- Applications of mass spectrometry in the routine diagnostic medical laboratory – a status report 2025Journal of Chromatography B · 2026
- Workflow for Assessing Mass Spectrometer Performance in Shotgun LipidomicsbioRxiv · 2026
- Preliminary analytical evaluation of the fully automated Cobas i601 LC-MS/MS system for steroid hormones and vitamin D in a clinical laboratory.Clinica chimica acta; international journal of clinical chemistry · 2026
- A-280 Single Center Evaluation of Alternative Assessment of Performance for Specific Immunology AnalytesClinical Chemistry · 2026