Workflow for Assessing Mass Spectrometer Performance in Shotgun Lipidomics

S. Radrezza, H. Moon, R. Ahrends, B. Bruegger, C. Coman, C. S. Ejsing, X. Han, M. Hoering,

Key points

From the paper's conclusion
  • We developed and validated the QC Shotgun workflow and web-accessible software that enable rapid testing, evaluation, documentation, and longitudinal monitoring of the performance of high-resolution mass spectrometers in quantitative shotgun lipidomics.

Abstract

Reproducible and testable analytical performance of mass spectrometers is critical for inter-laboratory concordance of quantitative lipidomics analyses. However, common instrument readout or analyses of arbitrarily chosen lipid mixtures do not lend themselves to direct comparison between instruments independently configured and operated in different laboratories under in-house settings. We developed and validated the QC Shotgun workflow and web-accessible software that enable rapid testing, evaluation, documentation, and longitudinal monitoring of the performance of high-resolution mass spectrometers in quantitative shotgun lipidomics. Completed in less than 5 minutes, the QC Shotgun test reports 8 relative metric indices that facilitate inter-laboratory comparison and longitudinal assessment of instrument performance and identify deficiencies that are not apparent from instrument readouts. Systematic documentation of instrument status can improve inter-laboratory consistency and interpretation confidence of lipidomics datasets.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers

Workflow for Assessing Mass Spectrometer Performance in Shotgun Lipidomics — Science Skim