Mitigating the effects of breast and excitation-source misalignment in low-frequency breast cancer detection.
Ghazaleh Tashtarian, Ahad Tavakoli, Abdolali Abdipour, Ali M Almuhlafi, Omar M Ramahi
Department of Electrical Engineering, Amirkabir University of Technology (Tehran Polytechnic), Tehran, Iran. · Department of Electrical Engineering, King Saud University, Riyadh, Saudi Arabia.
Key points
From the paper's conclusion- The proposed alignment method effectively mitigates excitation-source and breast misalignment in low-frequency breast imaging, enabling non-ionizing tumor detection with accurate 2-D localization.
Abstract
OBJECTIVES: Accurate breast-tumor detection requires precise positioning of both the excitation source and the breast, as motion or misalignment can degrade imaging performance. This work presents a novel technique to compensate for this effect and to reliably identify tumor presence and 2-D localization in breast impressions. METHODS: A numerical alignment approach was developed to estimate source-breast misalignment via cross-correlation and compensate for the resulting displacement using Fourier-domain phase compensation. An electrically small 200-MHz loop antenna was used as a non-ionizing excitation source to provide adequate penetration. In addition, a method was developed to generate MRI-derived contralateral breast phantoms based on enantiomorphic breast anatomy, thereby enhancing simulation realism. RESULTS: Comprehensive simulations on a realistic heterogeneously dense breast model demonstrate that the proposed technique consistently detects tumors of varying sizes, depths, and locations under excitation-source and breast misalignment in the presence of AWGN with an SNR of 30 dB. The robustness of the method was further assessed under different noise levels. Quantitative positioning-error analysis confirms high localization accuracy, yielding an average positioning-error reduction of 85.75 % across the evaluated simulation scenarios. CONCLUSIONS: The proposed alignment method effectively mitigates excitation-source and breast misalignment in low-frequency breast imaging, enabling non-ionizing tumor detection with accurate 2-D localization.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Dosimetric Vulnerability of Radiotherapy Techniques Under Non-Rigid Arm-Position-Induced Deformations in Breast Cancer: A Deformable Phantom StudyRadiation · 2026
- Sensitivity and Identifiability of Tumor-Related Changes in Longitudinal Microwave Breast ImagingarXiv · 2026
- A Projector-Based Augmented-Reality Localization Device for Breast-Conserving Surgery: Phantom Pipeline and End-to-End Error Budget.Bioengineering (Basel, Switzerland) · 2026
- Fourier Multiscale Analysis of Bilateral Breast Asymmetry Associated with Breast CancermedRxiv · 2026
- Recoverability-guided reduced-target inversion for microwave imaging: a synthetic breast-imaging studyBiomedical Physics & Engineering Express · 2026
- Supine Magnetic Resonance Imaging in Breast-Conserving Surgery Planning: A Systematic Review of Margin and Re-excision Outcomes.Clinical breast cancer · 2026