A Projector-Based Augmented-Reality Localization Device for Breast-Conserving Surgery: Phantom Pipeline and End-to-End Error Budget.
Xiaonan Gong, Jiang Heng, Hao Feng, Suzhan Zhang
Department of Breast Surgery, Second Affiliated Hospital, Zhejiang University School of Medicine, 88 Jiefang Road, Hangzhou 310009, China. · Cloud Valley, No. 1008 Dengcai Street, Xihu District, Hangzhou 310030, China.
Key points
From the paper's conclusion- We instead report a component-based end-to-end error budget obtained as the root-sum-square of three independently measured component errors: 2.7 mm (RSS planning estimate, not a directly measured end-to-end accuracy) -rigid CT 0.30 mm, ProCam physical projection 2.50 mm, ICP 0.92 mm.
Abstract
Breast-conserving surgery (BCS) requires accurate intraoperative localization of the tumor, which moves between the prone diagnostic and supine surgical positions. We present a projector-based augmented-reality (AR) localization system that projects a crosshair directly onto the patient's skin at the predicted tumor location. The pipeline combines rigid registration of ceramic fiducials between prone and supine CT (tumor error 0.30 mm on a real phantom), a SOFA finite-element simulation of gravity-driven prone-to-supine soft-tissue deformation (mean bead displacement 24.7 mm on a synthetic mesh; not validated on a deformable phantom), two-stage ICP surface alignment (RMSE 0.92 mm on a simulated camera; not yet validated on real D415 acquisitions), and projector-camera (ProCam) extrinsic calibration driving the crosshair projection. Because the tumor and sub-surface fiducials are not visible in the live RGB-D scene, a direct non-circular in vivo target registration error (TRE) on the tumor cannot be measured in the present phantom configuration. We instead report a component-based end-to-end error budget obtained as the root-sum-square of three independently measured component errors: 2.7 mm (RSS planning estimate, not a directly measured end-to-end accuracy) -rigid CT 0.30 mm, ProCam physical projection 2.50 mm, ICP 0.92 mm. The ProCam term is largest but sub-3 mm. The Gray-code reprojection residual (~11 mm in 15-pose recalibration; 8.62 mm in the deployed 4-pose calibration) overestimates the actual projection error by ~4× (resp. ~3.4×): least-squares calibration averages the consumer-DLP vertical quantization (16-level Gray-code decode on XPR-DLP) into an accurate rigid transform, so the per-observation residual reflects observation quantization, not projection accuracy. The quantization's measurable cost is projection variance (max 5.5 mm), which motivates phase-shifting structured light. Boundary conditions on phantom rigidity, sub-surface fiducials, and simulated-camera ICP are stated explicitly.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Study Protocol for Assessment of the Feasibility of an Intraoperative Localization Technique Using Augmented Reality in Breast-Conserving Surgery for Non-Palpable Cancer Lesions: A Feasibility Study.Health science reports · 2026
- Novel Intraoperative Augmented Reality-Guided Specimen Localization in Chest Wall Oncologic Surgery: A Case Report.Journal of medical extended reality · 2026
- Augmented Reality-based Guidance with Deformable Registration in Head and Neck Tumor Resection.Medical image computing and computer-assisted intervention : MICCAI ... International Conference on Medical Image Computing and Computer-Assisted Intervention · 2026
- Physics-informed depth-enhanced surface tracking for cumulative dose reconstruction in breast radiotherapy: An in silico feasibility study.Journal of applied clinical medical physics · 2026
- Dynamic infrared thermography and computed tomography angiography fusion for DIEP flap perforator selection: a novel measurement systemarXiv · 2026
- Manufacturing and effectiveness evaluation of effective breast suit for breast radiotherapyOSF Preprints (OSF Preprints) · 2026