Repeatability and Reliability of Corneal Tomography in Keratoconus Patients with Corneal Allogenic Intrastromal Ring Segments.
Ugur Tunc, Dilan Colak, Burcu Yakut, Fatma FeyzaNur Keskin Perk, Metin Suleymanzade, Sepideh Lotfi, Aylin Kılıç
Department of Ophthalmology, Istanbul Medipol University, Faculty of Medicine, Istanbul, Türkiye. · Department of Ophthalmology, Beyoglu Eye Training ad Research Hospital, Istanbul, Türkiye.
Key points
From the paper's conclusion- The implantation of CAIRS alters the shape of the cornea and decreases the repeatability of several tomography-derived metrics, especially posterior corneal indices and Kmax. Although overall reliability remains high, clinicians should cautiously interpret small postoperative changes in monitoring these parameters. Pachymetric measurements maintain excellent precision and can be reliably used for postoperative follow-up and progression assessment after CAIRS surgery.
Abstract
OBJECTIVES: To assess the repeatability and reliability of anterior and posterior corneal measurements in keratoconus patients who underwent corneal allogenic intrastromal ring segment (CAIRS) implantation using a Scheimpflug-based tomography system. METHODS: This prospective observational study included 42 eyes: 20 keratoconic eyes that had undergone CAIRS implantation and 22 healthy controls. All participants were imaged with the Pentacam HR (Oculus Optikgeräte GmbH, Germany). Each eye underwent three consecutive scans during a single session. Repeatability was evaluated using within-subject standard deviation (Sw), repeatability index (RI), and tolerance index (TI), and reliability was assessed using intraclass correlation coefficients (ICC). A TI>0.36 was considered indicative of reduced repeatability. RESULTS: Anterior corneal curvature parameters (K1 and K2) and pachymetric indices demonstrated excellent reliability (ICC>0.90) and acceptable repeatability (TI<0.36). In contrast, CAIRS-implanted eyes exhibited greater variability in Kmax, Kmean, anterior elevation, and several posterior corneal metrics, with TI values exceeding 0.36. Despite having high ICC values, posterior elevation showed the greatest variability. Pachymetry at the corneal vertex and the thinnest corneal thickness remained highly reproducible (TI 0.08 and 0.18, respectively), suggesting strong measurement accuracy. CONCLUSION: The implantation of CAIRS alters the shape of the cornea and decreases the repeatability of several tomography-derived metrics, especially posterior corneal indices and Kmax. Although overall reliability remains high, clinicians should cautiously interpret small postoperative changes in monitoring these parameters. Pachymetric measurements maintain excellent precision and can be reliably used for postoperative follow-up and progression assessment after CAIRS surgery.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Repeatability of interpolated corneal tomography maps in healthy and keratoconus corneas before and after crosslinking.Acta ophthalmologica · 2026
- Visual, Tomographic, and Biomechanical Outcomes of CAIRS with Accelerated Cross-Linking in Moderate to Advanced Keratoconus: A 6-Month Retrospective Case Series.Clinical ophthalmology (Auckland, N.Z.) · 2026
- Clinical and Tomographic Outcomes of Long Arc-Length Intrastromal Corneal Ring Segment Implantation in Central Keratoconus.Clinical ophthalmology (Auckland, N.Z.) · 2026
- Is Kmax a Reliable Endpoint for Evaluating Outcomes after CAIRS?Ophthalmology and therapy · 2026
- Anterior Stromal Melt Following Implantation of a Large Size Corneal Allogenic Intrastromal Ring Segment (CAIRS).Journal of refractive surgery (Thorofare, N.J. : 1995) · 2026
- Agreement between Sirius Scheimpflug-Placido tomography and Anterion swept-source OCT in keratoconic eyes stratified by prior corneal cross-linking: a retrospective cross-sectional agreement study.International ophthalmology · 2026