Vivaray Hb pro non-invasive hemoglobin device: A prospective diagnostic accuracy study measuring agreement with a calibrated blood cell counter
Devesh Kumar, Sunita Kapoor, Ambanna Gowda, Dipti Gupta, Harsh Mittal, S Sood
IMU University
Key points
From the paper's conclusion- The Vivaray Hb pro can function as a useful non-invsive screening and point-of-care assessment device for measuring hemoglobin, particularly in settings where rapid, painless and repeatable measurements are desirable.
Abstract
BACKGROUND: Non-invasive hemoglobin measurement offers a painless and rapid alternative to conventional blood-based testing. The Vivaray Hb pro is a handheld photoplethysmography-based device designed for point-of-care hemoglobin assessment without blood sampling. We evaluated the clinical performance of the Vivaray Hb pro by comparing device-generated hemoglobin values with those obtained from a calibrated laboratory blood cell counter. METHODS: In this cross-sectional, non-randomized clinical performance study, participants aged ≥8 years were prospectively recruited. Hemoglobin was measured non-invasively using the Vivaray Hb pro and compared with venous blood samples analyzed on a calibrated Coulter counter. Agreement between methods was assessed using Bland-Altman analysis, including regression-based evaluation for proportional bias. Mean difference, mean absolute error (MAE) and proportions of measurements within tolerance limits were also calculated. Complete paired measurements were available for 763 individuals. RESULTS: Bland-Altman analysis returned a mean difference of -0.40 g/dL. The bias was hemoglobin-dependent, being higher than the true value at low hemoglobin levels and underestimation at higher levels. Regression-based analysis showed proportional bias (β1 = -0.178), indicating decreasing difference with increasing hemoglobin concentration. The MAE was 1.5 g/dL but was lower (1.2) in the clinically predominant ranges of 8.1-13 g/dL. CONCLUSION: The Vivaray Hb pro can function as a useful non-invsive screening and point-of-care assessment device for measuring hemoglobin, particularly in settings where rapid, painless and repeatable measurements are desirable.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Research on Intraoperative Non-Invasive Hemoglobin Detection Technology Based on Four-Wavelength PhotoplethysmographyBiosensors · 2026
- Noninvasive hemoglobin measurement in the emergency department: agreement, anemia detection, and the effect of perfusion index.BMC emergency medicine · 2026
- Validation of Non-Invasive Hemoglobin Measurement and Its Utility for Anemia Screening in Pregnancy: A Cross-Sectional Diagnostic Accuracy Study.Diagnostics (Basel, Switzerland) · 2026
- Accuracy of minimally invasive hemoglobin monitoring in the perioperative period: transcutaneous hemoglobin monitoring and point-of-care testing devices: prospective observational study.Journal of clinical monitoring and computing · 2026
- Validation of Point-of-care Testing Device HemoCue® Glucose 201RT in Blood Glucose Estimation.Indian journal of community medicine : official publication of Indian Association of Preventive & Social Medicine · 2026
- GlucoNeo Check in the community: feasibility of photoplethysmography-based non-invasive glucose screening.International journal of public health · 2026