BMC emergency medicine

Noninvasive hemoglobin measurement in the emergency department: agreement, anemia detection, and the effect of perfusion index.

Buğra İlhan, Yusuf Semih Koçak, Eren Talha Aras, Ali Murtaza Koçak, Cansu Çiçek, Oğuz Eroğlu, Turgut Deniz

Department of Emergency Medicine, Kırıkkale University Faculty of Medicine, Kırıkkale, Türkiye. bugra_ilhan@yahoo.com. · Department of Emergency Medicine, Kırıkkale University Faculty of Medicine, Kırıkkale, Türkiye.

Observational · n=861

Key points

From the paper's conclusion
  • Single SpHb values do not agree sufficiently with laboratory hemoglobin to replace it. Discrimination for severe anemia was high, but the proposed cut-offs are candidate thresholds requiring independent prospective validation.

Abstract

BACKGROUND: Noninvasive hemoglobin (SpHb) measurement could provide rapid results in the emergency department (ED), but its accuracy remains uncertain. We evaluated the agreement of two pulse CO-oximeters with laboratory hemoglobin, their performance for detecting anemia, the effect of perfusion index (PI), and the threshold dependence of performance. METHODS: In this prospective diagnostic accuracy study conducted in a tertiary university hospital ED (December 2024-January 2026), consecutive adult patients undergoing hemoglobin testing had SpHb measured with Rad-97 and Rad-67 pulse CO-oximeters; a hematology analyzer served as the reference standard. We used Bland-Altman, receiver operating characteristic (ROC), and decision curve analyses and varied the target threshold from 7.0 to 12.0 g/dL. RESULTS: Of 861 patients, 313 (36.4%) had anemia. Bias (laboratory - device) was + 0.88 g/dL for Rad-97 (limits of agreement - 2.05 to 3.81) and + 0.13 g/dL for Rad-67 (- 3.19 to 3.46). Both devices showed proportional bias and overestimated hemoglobin below 8 g/dL (bias - 1.62 and - 1.67 g/dL). The area under the ROC curve (AUC) for anemia was 0.855 and 0.832 (difference 0.023; 95% confidence interval [CI] - 0.005 to 0.051; p = 0.109). For severe anemia (< 8 g/dL; n = 30), the Rad-97 AUC was 0.974, and a cut-off of ≤ 10.2 g/dL yielded 96.7% sensitivity (95% CI 82.8-99.9) and 88.1% specificity; optimism-corrected sensitivity (94.1%; 95% CI 86.3-97.4) fell below the 95% criterion used to select this cut-off. PI had no clinically meaningful effect on accuracy. Measurement took a median of 2.0 and 0.83 min versus 21 min for the laboratory. CONCLUSIONS: Single SpHb values do not agree sufficiently with laboratory hemoglobin to replace it. Discrimination for severe anemia was high, but the proposed cut-offs are candidate thresholds requiring independent prospective validation.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers