Noninvasive hemoglobin measurement in the emergency department: agreement, anemia detection, and the effect of perfusion index.
Buğra İlhan, Yusuf Semih Koçak, Eren Talha Aras, Ali Murtaza Koçak, Cansu Çiçek, Oğuz Eroğlu, Turgut Deniz
Department of Emergency Medicine, Kırıkkale University Faculty of Medicine, Kırıkkale, Türkiye. bugra_ilhan@yahoo.com. · Department of Emergency Medicine, Kırıkkale University Faculty of Medicine, Kırıkkale, Türkiye.
Key points
From the paper's conclusion- Single SpHb values do not agree sufficiently with laboratory hemoglobin to replace it. Discrimination for severe anemia was high, but the proposed cut-offs are candidate thresholds requiring independent prospective validation.
Abstract
BACKGROUND: Noninvasive hemoglobin (SpHb) measurement could provide rapid results in the emergency department (ED), but its accuracy remains uncertain. We evaluated the agreement of two pulse CO-oximeters with laboratory hemoglobin, their performance for detecting anemia, the effect of perfusion index (PI), and the threshold dependence of performance. METHODS: In this prospective diagnostic accuracy study conducted in a tertiary university hospital ED (December 2024-January 2026), consecutive adult patients undergoing hemoglobin testing had SpHb measured with Rad-97 and Rad-67 pulse CO-oximeters; a hematology analyzer served as the reference standard. We used Bland-Altman, receiver operating characteristic (ROC), and decision curve analyses and varied the target threshold from 7.0 to 12.0 g/dL. RESULTS: Of 861 patients, 313 (36.4%) had anemia. Bias (laboratory - device) was + 0.88 g/dL for Rad-97 (limits of agreement - 2.05 to 3.81) and + 0.13 g/dL for Rad-67 (- 3.19 to 3.46). Both devices showed proportional bias and overestimated hemoglobin below 8 g/dL (bias - 1.62 and - 1.67 g/dL). The area under the ROC curve (AUC) for anemia was 0.855 and 0.832 (difference 0.023; 95% confidence interval [CI] - 0.005 to 0.051; p = 0.109). For severe anemia (< 8 g/dL; n = 30), the Rad-97 AUC was 0.974, and a cut-off of ≤ 10.2 g/dL yielded 96.7% sensitivity (95% CI 82.8-99.9) and 88.1% specificity; optimism-corrected sensitivity (94.1%; 95% CI 86.3-97.4) fell below the 95% criterion used to select this cut-off. PI had no clinically meaningful effect on accuracy. Measurement took a median of 2.0 and 0.83 min versus 21 min for the laboratory. CONCLUSIONS: Single SpHb values do not agree sufficiently with laboratory hemoglobin to replace it. Discrimination for severe anemia was high, but the proposed cut-offs are candidate thresholds requiring independent prospective validation.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Validation of Non-Invasive Hemoglobin Measurement and Its Utility for Anemia Screening in Pregnancy: A Cross-Sectional Diagnostic Accuracy Study.Diagnostics (Basel, Switzerland) · 2026
- Vivaray Hb pro non-invasive hemoglobin device: A prospective diagnostic accuracy study measuring agreement with a calibrated blood cell counterPLoS ONE · 2026
- Accuracy of minimally invasive hemoglobin monitoring in the perioperative period: transcutaneous hemoglobin monitoring and point-of-care testing devices: prospective observational study.Journal of clinical monitoring and computing · 2026
- Association between hemoglobin concentration and peripheral perfusion index in normovolemic patients: a prospective observational study.Journal of clinical monitoring and computing · 2026
- Research on Intraoperative Non-Invasive Hemoglobin Detection Technology Based on Four-Wavelength PhotoplethysmographyBiosensors · 2026
- Comparative Diagnostic Accuracy of Clinical Pallor at Different Anatomical Sites for Detecting Anemia in Pregnant Women Using HemoCue® as the Reference Standard.Cureus · 2026