Thermal Safety of Temperature Mapping During Trigeminal Radiofrequency Rhizotomy.
Michael D Staudt, Sanjeet S Grewal, Flora J Yang, Marshall T Holland, David P Darrow
Department of Neurological Surgery, University Hospitals Cleveland Medical Center, Case Western Reserve University School of Medicine, Cleveland, Ohio, USA. · Department of Neurological Surgery, Mayo Clinic, Jacksonville, Florida, USA.
Key points
From the paper's conclusion- Temperature mapping using multicontact cylindrical electrodes produced exposures several orders of magnitude below lesioning thresholds. These findings support a substantial thermal safety margin for temperature mapping during trigeminal RF rhizotomy.
Abstract
BACKGROUND AND OBJECTIVES: Radiofrequency (RF) rhizotomy of the Gasserian ganglion is an established treatment for trigeminal neuralgia that requires accurate localization before lesioning. Temperature-based interrogation has been proposed as a localization strategy, but its thermal safety margin has not been quantified. We quantified thermal injury risk of temperature mapping relative to definitive lesioning. METHODS: Thermocouple logs from 12 patients undergoing RF rhizotomy with a cylindrical platinum band electrode (0.8-mm diameter, 2.0-mm contact length) included 54 mapping and 19 definitive lesion episodes. Arrhenius injury dose was computed from temperature-time data. A reduced-order effective-medium model of the Gasserian ganglion, parameterized by cerebrospinal fluid fraction and distributed heat exchange, was calibrated to burn episodes and applied to mapping episodes. Sensitivity analyses evaluated 2 Arrhenius parameter sets and ±2°C thermocouple bias. RESULTS: The Arrhenius damage integral (Ω; Ω ≥ 1 indicating predicted irreversible injury) at the thermocouple was far lower during mapping than during definitive lesioning. Under neural-relevant parameters, median mapping Ω was 3.73 × 10-4 (IQR 4.16 × 10-5-1.65 × 10-3; maximum 9.92 × 10-3) vs 1.85 × 103 (IQR 5.90 × 101-3.36 × 103) during burns. Predicted burn lesion radii were 1.8 to 2.2 mm, while mapping radius was 0 mm across all episodes. Mapping remained subthreshold under canonical RF parameters and ±2°C bias. CONCLUSION: Temperature mapping using multicontact cylindrical electrodes produced exposures several orders of magnitude below lesioning thresholds. These findings support a substantial thermal safety margin for temperature mapping during trigeminal RF rhizotomy.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- RF-EMF response modeling of implantable medical electrodes under 1.5T and 3T MRI exposure using a lumped element model approach.Physics in medicine and biology · 2026
- Assessment of the accuracy and feasibility of CT fusion with real-time US guidance for percutaneous radiofrequency trigeminal rhizotomy.European journal of radiology · 2026
- Imaging Informatics for Thermal Ablation: CT Thermometry-Based Tissue Injury Mapping.IEEE journal of biomedical and health informatics · 2026
- Assumption-bounded reduced electro-thermal modeling of lesion formation in radiofrequency cardiac ablation: power-time allocation, thermal latency, perfusion, and temperature limitation.Computer methods in biomechanics and biomedical engineering · 2026
- Radiofrequency Ablation Versus Cryoablation of the Posterior Nasal Nerve for Chronic Rhinitis: A Meta-Analysis and Systematic Review.American journal of rhinology & allergy · 2026
- Ultrasound-guided hydrodissection combined with radiofrequency for refractory occipital neuralgia: clinical outcomes, durability, and factors associated with repeat intervention.Frontiers in neurology · 2026