Operative neurosurgery (Hagerstown, Md.)

Thermal Safety of Temperature Mapping During Trigeminal Radiofrequency Rhizotomy.

Michael D Staudt, Sanjeet S Grewal, Flora J Yang, Marshall T Holland, David P Darrow

Department of Neurological Surgery, University Hospitals Cleveland Medical Center, Case Western Reserve University School of Medicine, Cleveland, Ohio, USA. · Department of Neurological Surgery, Mayo Clinic, Jacksonville, Florida, USA.

Key points

From the paper's conclusion
  • Temperature mapping using multicontact cylindrical electrodes produced exposures several orders of magnitude below lesioning thresholds. These findings support a substantial thermal safety margin for temperature mapping during trigeminal RF rhizotomy.

Abstract

BACKGROUND AND OBJECTIVES: Radiofrequency (RF) rhizotomy of the Gasserian ganglion is an established treatment for trigeminal neuralgia that requires accurate localization before lesioning. Temperature-based interrogation has been proposed as a localization strategy, but its thermal safety margin has not been quantified. We quantified thermal injury risk of temperature mapping relative to definitive lesioning. METHODS: Thermocouple logs from 12 patients undergoing RF rhizotomy with a cylindrical platinum band electrode (0.8-mm diameter, 2.0-mm contact length) included 54 mapping and 19 definitive lesion episodes. Arrhenius injury dose was computed from temperature-time data. A reduced-order effective-medium model of the Gasserian ganglion, parameterized by cerebrospinal fluid fraction and distributed heat exchange, was calibrated to burn episodes and applied to mapping episodes. Sensitivity analyses evaluated 2 Arrhenius parameter sets and ±2°C thermocouple bias. RESULTS: The Arrhenius damage integral (Ω; Ω ≥ 1 indicating predicted irreversible injury) at the thermocouple was far lower during mapping than during definitive lesioning. Under neural-relevant parameters, median mapping Ω was 3.73 × 10-4 (IQR 4.16 × 10-5-1.65 × 10-3; maximum 9.92 × 10-3) vs 1.85 × 103 (IQR 5.90 × 101-3.36 × 103) during burns. Predicted burn lesion radii were 1.8 to 2.2 mm, while mapping radius was 0 mm across all episodes. Mapping remained subthreshold under canonical RF parameters and ±2°C bias. CONCLUSION: Temperature mapping using multicontact cylindrical electrodes produced exposures several orders of magnitude below lesioning thresholds. These findings support a substantial thermal safety margin for temperature mapping during trigeminal RF rhizotomy.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers