Assessment of the accuracy and feasibility of CT fusion with real-time US guidance for percutaneous radiofrequency trigeminal rhizotomy.
Chia-Hong Guo, Yung-Chi Hsu, Wei-Chen Liao, Hao-Lun Kao, Kung-Fu Li, Wei-Chou Chang, Yi-Chih Hsu
Department of Radiology, Tri-Service General Hospital, Taipei, Taiwan, ROC; National Defense Medical Center, Taiwan, ROC. · Department of Anesthesiology, Tri-Service General Hospital, Taipei, Taiwan, ROC; National Defense Medical Center, Taiwan, ROC.
Key points
From the paper's conclusion- CTF-USG demonstrated technical feasibility and quantifiable spatial accuracy for PRF-TR in TN. Objective spatial measurements and high interobserver reliability provide a reproducible framework for evaluating fusion-guided needle navigation. Larger comparative studies are warranted to determine its reproducibility and clinical benefit.
Abstract
PURPOSE: To quantitatively assess the spatial accuracy and technical feasibility of computed tomography (CT) fusion with real-time ultrasound (US) guidance (CTF-USG) for percutaneous radiofrequency trigeminal rhizotomy (PRF-TR) in patients with trigeminal neuralgia (TN). METHODS: This prospective observational study included consecutive CTF-USG procedures performed between July 2024 and December 2025. Preprocedural CT-US registration accuracy was quantified as synchronization error using predefined bony landmarks. Postprocedural needle-targeting accuracy was assessed by the three-dimensional distance between the final needle tip and the center of the intended region of interest (ROI) on CT. Technical success, radiofrequency sensory stimulation (RFSS), and identification of predefined bony landmarks and vessels were recorded. Interobserver reliability was assessed using intraclass correlation coefficients (ICCs) and Bland-Altman analysis. The association between synchronization error and needle tip-to-ROI distance was explored using Spearman rank correlation. RESULTS: Twenty-six target sites in 22 patients were analyzed. Mean CT-US synchronization error was 2.09 ± 1.67 mm, and mean needle tip-to-ROI distance was 2.65 ± 1.05 mm. Interobserver reliability was high (ICC[2,1] = 0.846 and 0.941), with close Bland-Altman agreement. First-attempt technical success and positive RFSS responses were achieved in all procedures. All predefined bony landmarks and vessels were identified, and no procedure-related complications occurred. No evident association was observed between synchronization error and needle tip-to-ROI distance (Spearman's ρ = 0.131, P = 0.522). CONCLUSION: CTF-USG demonstrated technical feasibility and quantifiable spatial accuracy for PRF-TR in TN. Objective spatial measurements and high interobserver reliability provide a reproducible framework for evaluating fusion-guided needle navigation. Larger comparative studies are warranted to determine its reproducibility and clinical benefit.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- OneRF radiofrequency ablation system for the minimally invasive treatment of trigeminal neuralgia: initial clinical experience.Pain management · 2026
- Multimodal MRI fusion and three-dimensional reconstruction for preoperative assessment of neurovascular compression in trigeminal neuralgia: A retrospective study.Neurosurgical review · 2026
- Thermal Safety of Temperature Mapping During Trigeminal Radiofrequency Rhizotomy.Operative neurosurgery (Hagerstown, Md.) · 2026
- Spatial accuracy and clinical feasibility of CT/US fusion imaging for hepatic interventions in dogs.BMC veterinary research · 2026
- Ultrasound-CT fusion versus ultrasound guidance for lumbar nerve root block in multilevel lumbar disc herniation: a retrospective comparative cohort study.BMC anesthesiology · 2026
- Comparative efficacy and safety of coblation and radiofrequency thermocoagulation in trigeminal neuralgia: a systematic review and meta-analysisBMC Neurology · 2026