Imaging Informatics for Thermal Ablation: CT Thermometry-Based Tissue Injury Mapping.
Tiansheng Liu, Xiaolong Wang, Jihong Chen, Jie Chen, Zhuting Fang, Bingwei He
Key points
From the paper's conclusion- Leakage-controlled calibration analysis indicated limited out-of-sample CT thermometry performance, whereas two-level validation supported the stability of the endpoint-temperature-to-injury mapping.
Abstract
Sparse CT acquisition during CT-guided thermal ablation limits direct estimation of cumulative ther mal injury, which depends on temperature history rather than endpoint temperature alone. This study developed an imaging-informatics pipeline for near real-time Arrhenius injury mapping from workflow-compatible endpoint CT thermometry. Endpoint temperature fields were derived from ROI-level HU-temperature calibration and deployed in baseline-referenced form. Protocol-specific normalized temperature-history templates reconstructed heating trajectories at 180 and 300 s, and precompiled endpoint temperature-to-logΩ mappings avoided voxel-wise temporal integration at runtime. Validation was limited to ex vivo liver specimens, using porcine tissue for model de velopment and validation and two bovine cases for ex ploratory cross-tissue evaluation. Across five indepen dent experiments, per-update latency was 793.7-1688.0 ms (1198.5±434.3 ms). The logΩ ≥ 0 delineation yielded mean absolute major- and minor-axis errors of 0.96±0.73 mm and 1.98 ± 1.37 mm, respectively, with Dice 0.9151 ± 0.0311 and HD952.07±0.27 mmrelative to gross specimen boundaries. Dice and HD95 quantify post-alignment two-dimensional contour agreement after scale normalization rather than absolute localization accuracy. Leakage-controlled calibration analysis indicated limited out-of-sample CT thermometry performance, whereas two-level validation supported the stability of the endpoint-temperature-to-injury mapping. These ex vivo results demonstrate protocol-specific feasibility and motivate independent thermometry validation, uncertainty propagation, and prospective in vivo evaluation.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- New analytical and hybrid heat transfer models for thermal ablation procedures validated by MRI thermometryInternational Journal of Hyperthermia · 2025
- Feasibility of prior‐informed domain‐based electrical impedance tomography for deep tissue thermometry: Phantom experiments and computed tomography‐derived numerical simulationsMedical Physics · 2026
- Deep Learning-Driven Computational Imaging for Noninvasive Monitoring System of Brain Temperature and Metabolism: A Hypothermia Validation for Acute Ischemic Stroke.MedComm · 2026
- Real-World Deployment and Operational Evaluation of an On-Premise Thermal IoMT System in an Intensive Care Unit.Journal of medical systems · 2026
- The Impacts of Atangana–Baleanu Fractional Derivative in Biological Tissues during ThermotherapyDOAJ (DOAJ: Directory of Open Access Journals) · 2026
- Dynamic Infrared Thermography for Quantitative Assessment of Peripheral Microcirculation: Models, Metrics, and Clinical TranslationPhysiological Measurement · 2026