Trajectory Metric Analysis in Stereotactic Robotic-Guided Stereoelectroencephalography: Fiducial-Based Intraoperative Computed Tomography Versus Laser Facial Scanning Registration Workflows.
Ryan R Song, Akshay Sharma, Nehaw Sarmey, Stephen Harasimchuk, Juan Bulacio, Richard Rammo, William Bingaman, Demitre Serletis
Cleveland Clinic Lerner College of Medicine of Case Western Reserve University, Cleveland, Ohio, USA. · Cleveland Clinic Epilepsy Center, Cleveland Clinic Foundation, Cleveland, Ohio, USA.
Key points
From the paper's conclusion- The intraoperative CT registration workflow, compared with laser facial scanning, provided significantly lower stereotactic error across all 3 error metrics. Laser facial scanning still offered an adequate error profile for SEEG. Depending on clinical necessity, these results validate intraoperative CT registration as an important step in accurate robotic-guided SEEG.
Abstract
BACKGROUND AND OBJECTIVES: Stereoelectroencephalography (SEEG) is an important methodology for invasive evaluation in medically refractory epilepsy patients. Traditionally, laser-based facial scanning has been used for stereotactic robotic registration at our center. More recently, we developed and validated a noninvasive, frame-based fiducial system integrated with intraoperative computed tomography (CT) registration, as part of our SEEG workflow. This study aims to compare the stereotactic accuracy in SEEG electrode placement using these 2 approaches. METHODS: Registration error, implantation accuracy metrics, and perioperative variables were collected for 56 refractory epilepsy patients undergoing SEEG between April 2022 and March 2024. Thirty-four patients underwent intraoperative CT registration, and 22 patients underwent laser facial scanning registration. Entry-point error, target-point error, and angular error metrics were measured for every electrode trajectory, across 1059 electrodes total. Wilcoxon rank sum tests were used to compare error measures between registration modalities. RESULTS: Overall mean age was 27.7 ± 11.2 years, with a mean of 28.3 ± 10.2 years in the laser group and 27.3 ± 12.0 years in the CT group. The average number of electrodes implanted was 19 ± 3 (intraoperative CT: 20 ± 3; laser: 19 ± 3). The average entry-point error was 1.52 ± 0.91 mm (intraoperative CT: 1.28 ± 0.83 mm; laser: 1.87 ± 0.90 mm; P < .0001). The average target-point error was 2.75 ± 1.48 mm (intraoperative CT: 2.65 ± 1.43 mm; laser: 2.89 ± 1.53 mm; P = .0050). The average angular error was 0.051 ± 0.038 radians (intraoperative CT: 0.045 ± 0.032 radians; laser: 0.060 ± 0.045 radians; P < .0001). CONCLUSION: The intraoperative CT registration workflow, compared with laser facial scanning, provided significantly lower stereotactic error across all 3 error metrics. Laser facial scanning still offered an adequate error profile for SEEG. Depending on clinical necessity, these results validate intraoperative CT registration as an important step in accurate robotic-guided SEEG.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Robotic Arm Alignment System for Electrode Implantation in Invasive Epilepsy Monitoring: Clinical Experience and Accuracy in a Consecutive Patient Series with Comparison to the Literature.Journal of clinical medicine · 2026
- Accuracy, Safety, and Workflow of Frameless Neuronavigation-Guided Stereoelectroencephalography in Drug-Resistant Epilepsy: A Single-Center Cohort Study.Operative neurosurgery (Hagerstown, Md.) · 2026
- Impact of entry trajectory angle-to-skull on electrode placement accuracy in robot-assisted stereoelectroencephalography.Epilepsy research · 2026
- A 2-Year Experience of Robot-Assisted Epilepsy Surgery Using a New Cranial Robotic System With Optical Tracking: Variables Influencing Accuracy and Complications.Neurosurgery practice · 2026
- SEEG Contact Detector: A 3D Slicer Extension for Automated Localisation of Intracranial Electrode ContactsmedRxiv · 2026
- Structured-light three-dimensional surface scanning with deep learning-assisted N-bar localizer detection for stereotactic registration: a technical feasibility study in deep brain stimulationScientific Reports · 2026