Impact of entry trajectory angle-to-skull on electrode placement accuracy in robot-assisted stereoelectroencephalography.
Chan Park, Jong Hyun Kim, Ji Hyun Kim, Ji Hye Shin, Haewon Roh
Department of Neurosurgery, Korea University Guro Hospital, Korea University College of Medicine, Seoul, Republic of Korea. · Department of Neurology, Korea University Guro Hospital, Korea University College of Medicine, Seoul, Republic of Korea.
Key points
From the paper's conclusion- Trajectory-to-skull angle is a key geometric determinant of electrode placement accuracy in robot-assisted SEEG. Quantitative angle thresholds identified in this study provide practical guidance for trajectory planning and may help minimize mechanical deviation, improve implantation safety, and enhance the reliability of seizure-network localization.
Abstract
OBJECTIVE: Despite the widespread use of robotic systems to improve stereotactic accuracy in stereoelectroencephalography (SEEG), targeting errors still occur. Several studies have suggested that the trajectory-to-skull angle is closely associated with targeting error among various influencing factors; however, quantitative analyses remain limited. Therefore, this study aimed to quantitatively evaluate the relationship between trajectory-to-skull angle and implantation accuracy and to establish practical cut-off values for surgical planning. METHODS: A total of 10 patients (103 electrodes) were included in this study and various clinical variables were analyzed to identify whether trajectory-to-skull angle has a significant relation with accuracy using multivariate linear regression. Then, receiver operating characteristic (ROC) curve analysis with Youden index optimization was applied to determine angle thresholds for achieving ≤ 3 mm accuracy. RESULTS: Multivariate analysis showed that both axial and coronal trajectory-to-skull angles were significantly associated with target error, with more tangential incidence angles corresponding to larger deviations (p < 0.01). Axial angle-to-skull was particularly associated with anteroposterior (Y-axis) error, whereas coronal angle-to-skull primarily predicted superoinferior (Z-axis) error. ROC curve analysis identified optimal angle thresholds of 11.8° for axial trajectories and 6.3° for coronal trajectories to predict targeting accuracy within 3 mm. When both axial and coronal angles were maintained within these cut-offs, the probability of achieving a target error ≤ 3 mm was 88.2%. CONCLUSION: Trajectory-to-skull angle is a key geometric determinant of electrode placement accuracy in robot-assisted SEEG. Quantitative angle thresholds identified in this study provide practical guidance for trajectory planning and may help minimize mechanical deviation, improve implantation safety, and enhance the reliability of seizure-network localization.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- A 2-Year Experience of Robot-Assisted Epilepsy Surgery Using a New Cranial Robotic System With Optical Tracking: Variables Influencing Accuracy and Complications.Neurosurgery practice · 2026
- Robotic Arm Alignment System for Electrode Implantation in Invasive Epilepsy Monitoring: Clinical Experience and Accuracy in a Consecutive Patient Series with Comparison to the Literature.Journal of clinical medicine · 2026
- Trajectory Metric Analysis in Stereotactic Robotic-Guided Stereoelectroencephalography: Fiducial-Based Intraoperative Computed Tomography Versus Laser Facial Scanning Registration Workflows.Operative neurosurgery (Hagerstown, Md.) · 2026
- Accuracy of robot-assisted cortical bone trajectory pedicle screw placement: CT-based retrospective cohort study.Acta orthopaedica Belgica · 2026
- Electroencephalography for Radiation-Free Orientation Estimation of Directional Deep Brain Stimulation LeadsbioRxiv · 2026
- Axial Convergence in Percutaneous Robotic Pedicle Screw Placement: Mixed-Effects Analysis of 1006 ScrewsSpine Open · 2026