Next-gen paediatric imaging: ultra-low dose chest photon-counting CT approaching chest radiograph doses.
Ione Limantoro, Kwinten Torfs, Joke Binst, Marijke Proesmans, Luc Breysem, Hilde Bosmans, Michael Aertsen
Department of Radiology, University Hospital of Leuven, Leuven, Belgium. ione.limantoro@uzleuven.be. · Department of Imaging and Pathology, Catholic University Leuven, Leuven, Belgium.
Key points
From the paper's conclusion- Our PCD ULDCT chest protocol, approaching chest radiograph doses, is promising for (follow-up of) focal abnormalities and follow-up imaging of known chest pathology in children.
Abstract
OBJECTIVE: Photon-counting detector (PCD) CT may reduce radiation dose while maintaining image quality, which is particularly promising for children due to their radiosensitivity. This feasibility study investigated whether an ultra-low dose CT (ULDCT) chest protocol on PCD CT, with a dose comparable to chest radiographs, could be developed to detect gross lung abnormalities in children. MATERIALS AND METHODS: This ethics-approved retrospective study consisted of two parts: (1) development of a ULDCT chest protocol using an anthropomorphic phantom and (2) qualitative evaluation of these scans performed in 24 consecutive patients (< 16 years) for various indications. Two paediatric radiologists assessed diagnostic quality for six chest pathologies using a 3-point Likert scale. Radiation dose metrics were extracted from a dose management system, and effective doses (ED) were calculated using age-appropriate conversion factors. Statistical analysis included mainly descriptive calculations. Doses were compared with routine chest radiography. RESULTS: The Lungman phantom was scanned using 5 different protocols. The optimal settings were identified as Sn100 kVp, 2 mAs and pitch 1.5, yielding an ED of 14 µSv. All paediatric scans were assessed as acceptable, but a preference for an increased dose protocol was given to rule out interstitial lung disease or bronchiectasis. The mean total ED from the localiser and ULDCT of the chest was 11.6 [IQR: 10.9-12.8] µSv compared to 8.2 [IQR: 6.2-13.6] µSv for standard chest radiography. CONCLUSION: Our PCD ULDCT chest protocol, approaching chest radiograph doses, is promising for (follow-up of) focal abnormalities and follow-up imaging of known chest pathology in children. KEY POINTS: Question Can the technique of photon-counting detector CT result in radiation dose reduction, while achieving adequate image quality? Findings An ultra-low dose CT chest protocol was developed with a comparable dose to a single chest radiograph. Relevance statement The ultra-low dose CT chest protocol is promising in the setting of (follow-up of known) focal chest abnormalities, long-term follow-up after congenital diaphragmatic hernia repair and follow-up of interstitial lung disease in children.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Pediatric Chest Photon-counting CT: Dose and Noise Behavior of Clinical Sn100 kVp and 90 kVp Protocols across Image Quality Levels - a Phantom Study.RoFo : Fortschritte auf dem Gebiete der Rontgenstrahlen und der Nuklearmedizin · 2026
- Radiation Dose and Image Quality of Photon-Counting Detector CT for Multiphasic Pre-Transplant Abdominal Imaging in Young Children Aged ≤36 Months: A Propensity-Matched Study.The British journal of radiology · 2026
- Photon-counting CT in pediatric patients: A multi-institutional evaluation of protocols and radiation exposure in 4,772 patientsDiagnostic and Interventional Imaging · 2025
- Quantitative and qualitative lung parameters in photon-counting CT: comparison of ultra-low-dose with standard low-dose protocols.European radiology · 2026
- Photon-Counting CT Enables Higher Image Quality at a Lower Radiation Dose Compared with a 256-Slice Energy-Integrating Detector CT in Pediatric Patients with Congenital Heart DiseaseDiagnostics · 2026
- Comparing 70 kV and 120 kV photon-counting CT protocols for pediatric cardiac imaging in children weighing 10-30 kg: Radiation dose and image quality.Journal of cardiovascular computed tomography · 2026