Quantitative and qualitative lung parameters in photon-counting CT: comparison of ultra-low-dose with standard low-dose protocols.
Jonas Kroschke, Lisa Jungblut, Oyunbileg von Stackelberg, Falko Ensle, Bjarne Kerber, Hans-Ulrich Kauczor, Thomas Frauenfelder, Mark Oliver Wielpütz,
Department of Diagnostic and Interventional Radiology, University Hospital Zurich (USZ), University of Zurich (UZH), Zurich, Switzerland. jonas.kroschke@usz.ch. · Department of Diagnostic and Interventional Radiology, University Hospital Zurich (USZ), University of Zurich (UZH), Zurich, Switzerland.
Key points
From the paper's conclusion- ULD-PCCT enables dose-efficient QCT but introduces limitations: reduced peripheral airway visibility/segmentation with differences in airway wall measurements, biased density-based emphysema quantification, and visual emphysema underestimation at ULD, supporting dose-aware interpretation and further validation of correction strategies.
Abstract
OBJECTIVES: Photon-counting CT (PCCT) enables dose reduction beyond energy-integrating CT, but ultra-low-dose acquisitions may impair assessment of airways and lung parenchyma. We evaluated the impact of ultra-low-dose (ULD) PCCT on quantitative and qualitative lung parameters. METHODS: In this prospective study, 101 patients underwent same-session low-dose (LD) and ULD chest CT for mixed indications (CTDIvol, 0.64 ± 0.15 vs. 0.11 ± 0.03 mGy). Automated QCT assessed noise index, airway segmentation, airway wall percentage, emphysema index (EI), mean lung density (MLD), and Percentile15 (Perc15) on soft (Br36u) and lung (Bl64u) kernels. Two readers scored image quality, airways, and emphysema extent using Likert scales. RESULTS: ULD increased noise compared with LD (Bl64u: +94.4%; Br36u: +32.8%) and reduced visual image quality (p < 0.001). Automated airway segmentation decreased (-41.4 mean no. of branches, p < 0.001), with fewer visible airway generations (p < 0.001). Visual bronchial wall thickening did not differ (R1 p = 0.89; R2 p = 0.20). Airway wall percentage showed kernel- and generation-dependent differences: whole-tree values did not differ on Br36u (LD-ULD -0.05%; p = 0.83) but were higher at ULD on Bl64u (LD-ULD -3.33%; p < 0.001), with greater differences peripherally. At ULD, MLD and Perc15 were lower than at LD by 12.5 and 19.9 HU, respectively (both p < 0.001), while EI was 6.2% higher. Visual emphysema assessment shifted toward lower severity at ULD (p < 0.001). CONCLUSION: ULD-PCCT enables dose-efficient QCT but introduces limitations: reduced peripheral airway visibility/segmentation with differences in airway wall measurements, biased density-based emphysema quantification, and visual emphysema underestimation at ULD, supporting dose-aware interpretation and further validation of correction strategies. KEY POINTS: Question To study the influence of ultra-low-dose photon-counting chest CT on quantitative and qualitative airway and emphysema assessment compared to low-dose CT. Findings Ultra-low-dose increased noise and reduced peripheral airway segmentation; airway wall metrics showed kernel- and generation-dependent differences, with biased density-based and visual emphysema assessment. Clinical relevance PCCT enables quantitative lung imaging at ultra-low dose, but dose reduction alters airway segmentation and density-based emphysema metrics. ULD-derived QCT therefore requires cautious interpretation and validation of dose-aware correction strategies before routine clinical use.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Ultra-Low-Dose Photon-counting Detector CT for Emphysema Assessment: A Head-to-Head Comparative Study with Low-Dose CTRadiology · 2026
- Next-gen paediatric imaging: ultra-low dose chest photon-counting CT approaching chest radiograph doses.European radiology experimental · 2026
- Low-dose vs standard-dose photon-counting CT for myopathy-associated interstitial lung disease: a prospective feasibility study.European radiology · 2026
- Photon-counting detector CT versus cone-beam CT for postoperative cochlear implant imaging.European journal of radiology · 2026
- Improvement of Lung Nodule Volumetric Accuracy with Photon-counting Computed Tomography Over Energy-integrating Computed Tomography in Low-dose ScreeningInvestigative Radiology · 2026
- Pulmonary nodule image quality and morphologic visualization with photon-counting detector CT: a scoping reviewEuropean Radiology Cardiovascular and Thoracic · 2026