AQbD-Driven RP-HPLC Method for Rapid Dissolution Monitoring of Amiodarone Hydrochloride Tablets Without Sample Dilution.
Ju-Hyun Yoon, Joo-Eun Kim
Department of Biopharmaceutical Chemistry, Kookmin University, Seoul 02707, Republic of Korea.
Key points
From the paper's conclusion- It is highly suitable for rapid routine quality control and high-precision dissolution testing of generic amiodarone pharmaceuticals in industrial settings.
Abstract
Background/Objectives: In this study, a novel reversed-phase high-performance liquid chromatography (RP-HPLC) method was established for the quality control of amiodarone hydrochloride, overcoming the limitations of the legacy compendial protocol. The conventional United States Pharmacopeia method requires a 10-fold sample dilution to avoid column overload and matrix interference from polysorbate 80. Attempting direct injection under conventional conditions severely compromises data reliability, leading to severe band broadening and peak distortion. Methods: To fairly evaluate the intrinsic separation capacity of the analytical systems, preliminary experiments were conducted using identical, undiluted dissolution samples. Through precise chromatographic redesign, the novel method eliminates the pre-treatment dilution step. Furthermore, the reliability, precision, and robustness of the proposed method were comprehensively validated in strict accordance with the International Council for Harmonisation Q2 guidelines. Results: Even upon direct injection of highly concentrated samples, the optimized system focuses broad peaks into sharp, symmetric, and narrow bands, increasing analytical sensitivity and theoretical plates. Furthermore, it maximizes intra-column separation efficiency, achieving complete baseline isolation of amiodarone from surfactant interferences within a retention time of approximately 6.6 min, thereby significantly improving analytical throughput. Conclusions: By eliminating cumbersome sample preparation, minimizing analysis time, and enhancing quantitative integrity, this validated RP-HPLC method offers an efficient, high-throughput standard analytical platform. It is highly suitable for rapid routine quality control and high-precision dissolution testing of generic amiodarone pharmaceuticals in industrial settings.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Comparative Evaluation and Improvement of the Analytical Method for Amiodarone Hydrochloride: Replacing the Pharmacopeial Method with a Validated RP-HPLC TechniqueApplied Sciences · 2026
- Determination of total coenzyme Q10 in seven types of dietary supplement matrices by high-performance liquid chromatography coupled with diode array detection.Journal of chromatography. A · 2026
- Development of a relative molar sensitivity-based HPLC method for quantifying acetaminophen, two beta-lactams, amiodarone, and its metabolite.Journal of pharmaceutical health care and sciences · 2026
- Central Composite Design Approach Stability-Indicating RP-HPLC Analysis for Simultaneous Estimation of Monoclonal Antibodies in Bulk and Combined Dosage Form.Assay and drug development technologies · 2026
- Development and Validation of a Quality by Design‐Driven RP‐HPLC Assay Method for the Simultaneous Estimation of Lenalidomide and Sonrotoclax in Bulk Drug Substance, with a Six‐Edged Sustainability AssessmentSEPARATION SCIENCE PLUS · 2026
- QbD-driven RP-HPLC method for the simultaneous analysis of dihydropyridines calcium channel blockers in pharmaceuticalsBMC Chemistry · 2025