Stochastic evaluation of the reliability of relative molar sensitivity for quantitative HPLC analysis using a chemometric tool based on the FUMI theory.
Kohei Matsumoto, Yuzuru Hayashi, Akira Kotani
Faculty of Pharmacy, Osaka Medical and Pharmaceutical University, 4-20-1 Nasahara, Takatsuki, Osaka, 569-1094, Japan. · Institute of FUMI Theory, 3-3-15 Inaridai, Sakura, Chiba, 285-0864, Japan.
Key points
From the paper's conclusion- In conclusion, our results demonstrate that the reliability of RMSs in quantitative HPLC analysis can be evaluated from a one-point calibration curve and a single chromatogram with a chemometric tool based on the FUMI theory.
Abstract
A single-reference HPLC method based on relative molar sensitivity (RMS) has attracted increasing attention as a quantitative approach that does not require analyte standards. RMS is defined as the ratio of the slope of calibration curves for an analyte and a reference substance. Therefore, the reliability of RMS is evaluated based on the confidence intervals of the calibration curves, which are conventionally estimated statistically from repetitive measurements at multiple concentrations. To reduce the resources required to evaluate the reliability of RMS, we propose a chemometric tool based on the function of mutual information (FUMI) theory, which enables the stochastic estimation of the standard deviation of area created by baseline noise from a single chromatogram. In this study, an HPLC method for determining salicylic acid and phenol in compound salicylic acid spirit was used as an example, with theophylline as the reference substance. The 95% confidence interval of the RMS for salicylic acid estimated using the chemometric tool (0.1559 - 0.1572) was within that obtained using the conventional method (0.1532 - 0.1610), indicating that the FUMI theory is applicable to evaluating the reliability of RMS. Similar results were obtained for the RMS for phenol. In analyses of real samples, the salicylic acid and phenol contents determined using the RMSs showed good agreement with those determined using the absolute calibration method. In conclusion, our results demonstrate that the reliability of RMSs in quantitative HPLC analysis can be evaluated from a one-point calibration curve and a single chromatogram with a chemometric tool based on the FUMI theory. This approach enables the reliability of RMS to be evaluated with substantially fewer measurements than the conventional method.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Development of a relative molar sensitivity-based HPLC method for quantifying acetaminophen, two beta-lactams, amiodarone, and its metabolite.Journal of pharmaceutical health care and sciences · 2026
- Metrological Evaluation of Metopimazine HPLC Assay: ISO-GUM and Monte Carlo Simulation ApproachesPharmaceutics · 2025
- MSlineaR: An R Package Assessing Linear Behavior to Improve Quality Assurance and Statistical Robustness in Untargeted Metabolomics.Analytical chemistry · 2026
- Development and Validation of an RP‐HPLC Method for the Rapid and Sensitive Quantification of Everolimus in Pharmaceutical and Medical Device ApplicationsSEPARATION SCIENCE PLUS · 2026
- A Feasible 1H-NMR Quantification Method of Indicator Compounds in Microwave-Processed Reynoutria multiflora Radix and Its Decoction.International journal of analytical chemistry · 2026
- Determination of Naproxen Impurities' Relative Response Factors by qNMR-HPLC Using DMSO-d6 Solvent as an Internal Calibrant.Molecules (Basel, Switzerland) · 2026