BMC oral health

Study on the stability of the end-effector in a zygomatic implant robot.

Xin Zhang, ChunHao Yang, Que Feng, Jing Xu, Jing Cao, ChangJian Li, Li Peng Jiang, Yue Zhang,

Department of Stomatology, the Fourth Medical Center of PLA General Hospital, No. 51 Fucheng Road, Beijing, Haidian District, 10048, China. · Department of Radiation Oncology, The First Affiliated Hospital of Jinzhou Medical University, Jinzhou, 121000, China.

Key points

From the paper's conclusion
  • The results showed that the dynamic eccentricity of zygomatic drills was greater than that of alveolar ridge drills when the end effector of the robotic arm was equipped with either a straight or an angled handpiece. Furthermore, when using a zygomatic drill under different robotic postures, the dynamic eccentricity was greater with an angled handpiece compared to a straight handpiece. This study provides a reference for the clinical application of angled handpieces as end effectors in zygomatic implant robots, especially when using extra-long drills.

Abstract

OBJECTIVE: To assess the impact of a zygomatic implant robotic arm on the dynamic eccentricity of an extra-long drill when affixed with either a straight or angled implant handpiece in various postural positions. METHODS: 1. In the same postural position, the zygomatic implant robot used straight and angled handpieces as end-effectors, each with twenty handpieces per group (n = 20). It tested the dynamic eccentricity of the same zygomatic drill and the alveolar ridge drill. 2. Using straight and angled handpieces (n = 20) as end-effectors for mounting the zygomatic drill, the zygomatic implant robot's arm was extended to its longest and shortest postural positions, as well as tested in a fixed non-robotic arm, to measure the dynamic eccentricity. The results were imported into SPSS for nonparametric Mann-Whitney U and Kruskal-Wallis H tests. RESULTS: The dynamic eccentricity of zygomatic drills was significantly greater than that of alveolar ridge drills when mounted on the same implant handpiece (p < 0.001). When the same drill was used, angled handpieces exhibited significantly higher dynamic eccentricity than straight handpieces (p < 0.001). In addition, the dynamic eccentricity of the implant handpiece increased significantly when the robotic arm was positioned at its longest posture compared with the shortest posture position and the non-robotic fixation condition (p < 0.001). No significant difference was observed between the shortest robotic posture and the non-robotic fixation condition (p > 0.05). CONCLUSION: The results showed that the dynamic eccentricity of zygomatic drills was greater than that of alveolar ridge drills when the end effector of the robotic arm was equipped with either a straight or an angled handpiece. Furthermore, when using a zygomatic drill under different robotic postures, the dynamic eccentricity was greater with an angled handpiece compared to a straight handpiece. This study provides a reference for the clinical application of angled handpieces as end effectors in zygomatic implant robots, especially when using extra-long drills.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers

Study on the stability of the end-effector in a zygomatic implant robot — Science Skim