Effect of occlusal splint design and thickness on temporomandibular joint loading: a finite element study
Başak Topdağı, Elifnur Guzelce Sultanoglu, Alanur Çiftçi Şişman, Hüsna Aktürk, Kübra Karaosmanoğlu
Sağlık Bilimleri Üniversitesi · Istinye University
Key points
From the paper's conclusion- This study compared the stress and strain distributions within the temporomandibular joint (TMJ) and adjacent tissues produced by Michigan (stabilization) and non-permissive splints of 3-mm and 5-mm thicknesses using finite element analysis.
Abstract
OBJECTIVES: Occlusal splints are a mainstay in the management of temporomandibular disorders, but the mechanical influence of design and thickness variations on joint and dental structures remains unclear. This study compared the stress and strain distributions within the temporomandibular joint (TMJ) and adjacent tissues produced by Michigan (stabilization) and non-permissive splints of 3-mm and 5-mm thicknesses using finite element analysis. MATERIALS AND METHODS: Three-dimensional models of the maxilla, mandible, teeth, periodontal ligament, and articular disc were reconstructed from the Visible Human Project dataset. Michigan and non-permissive splints were modeled with occlusal thicknesses of 3 mm and 5 mm. Functional muscle forces were applied to simulate mandibular movements, and linear static analyses were performed in OptiStruct under defined boundary conditions. RESULTS: The splint-free model showed the highest stress levels in all structures. Non-permissive splints significantly reduced mandibular stress (32.3 MPa) but increased loading on the periodontal ligament. Michigan splints resulted in lower periodontal ligament stresses and a more even stress distribution within the articular disc. Increasing splint thickness improved stress reduction only in the non-permissive model. CONCLUSIONS: Occlusal splints lowered peak stresses in the TMJ and related structures. Non-permissive splints provided greater joint protection, while Michigan splints achieved a more balanced distribution of occlusal forces. CLINICAL RELEVANCE: Splint design and thickness affect how loads are transferred through the temporomandibular system. Non-permissive splints, particularly at 5 mm, reduced joint stress but increased periodontal ligament loading. Splint selection should consider the main treatment goal, with attention to using the minimal effective thickness.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Patient-specific biomechanical evaluation of vertical TMJ splints in class II malocclusion with disc displacement: Influence of splint thickness and muscle fibre orientation.Proceedings of the Institution of Mechanical Engineers. Part H, Journal of engineering in medicine · 2026
- Biomechanical Evaluation of Different Posterior Mandibular Implant Configurations Under Bruxism: An FEA Study.Journal of functional biomaterials · 2026
- Effect of the Occlusal Splint Therapy on Mandibular Condylar Position in Patients With Myofascial Pain and an Average Skeletal Growth Pattern: A CBCT-Based Randomized Controlled Trial.Cureus · 2026
- Protrusive mandibular kinematics in anterior disc displacement: a CT-based 3D motion analysis before and after occlusal splint therapy.Journal of biomechanics · 2026
- Short-Term Clinical Outcomes of Ready-Made Soft Splints Versus Digitally Fabricated, Customized Hard Stabilization Splints in Temporomandibular Disorders: An Exploratory Randomized Study.Biomedicines · 2026
- An exploratory patient-specific finite element study of temporomandibular joint stress distribution following mandibular setback surgery in two distinct clinical scenarios.BMC oral health · 2026