Design and biomechanical investigation of a patient-specific total talar prosthesis for total talar replacement: A finite element analysis.
Md Qumar Tabrej, Amit Kumar, Jayanta Kumar Biswas
Department of Mechanical Engineering, NIT Patna, Bihar, India. · Department of Mechatronics and Automation Engineering, NIT Patna, Bihar, India.
Key points
From the paper's conclusion- The patient-specific TTP with screw fixation improves ankle stability in most loading directions, potentially compensating for ligament insufficiency. However, greater plantarflexion instability necessitates post-operative measures. These findings provide biomechanical evidence that fixation-enhanced TTP designs can improve therapeutic outcomes.
Abstract
BACKGROUND: Total talar replacement (TTR) is an emerging treatment for end-stage talar diseases, although concerns persist regarding ankle stability due to essential ligament transection. Further investigation is necessary regarding the biomechanical impact of patient-specific total talar prosthesis (TTP) on ankle stability and bone-implant mechanics. OBJECTIVE: The purpose of this study is designing a patient-specific TTP and assess its biomechanical performance on ankle stability and stress distribution under various loading conditions using finite element analysis. METHODS: A CT scan is used to create a 3D finite element model of the foot and ankle complex. The TTP made of A dual-material (UHMWPE talar body and titanium alloy cage) was created and fixed with five screws. Four loading conditions were simulated: inversion, eversion, dorsiflexion, and plantarflexion, with forces of 10 N, 50 N, 100 N, and 150 N. Talar tilt angles and von Mises stresses were examined for intact and implanted models. RESULTS: In comparison to the intact ankle, TTR significantly decreased the talar tilt angles in inversion (12.5%-16%), eversion (6.6%-14%), and dorsiflexion (33%-47%). On the other hand, plantarflexion produced a larger tilt (34%-44%). While bone stresses stayed below 5 MPa, maximum implant stresses ranged from 20 to 80 MPa. The validation of the model against cadaveric studies was reliable. CONCLUSION: The patient-specific TTP with screw fixation improves ankle stability in most loading directions, potentially compensating for ligament insufficiency. However, greater plantarflexion instability necessitates post-operative measures. These findings provide biomechanical evidence that fixation-enhanced TTP designs can improve therapeutic outcomes.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Position-Dependent Stability of an Unconstrained Total Talar Replacement and Its Suture Reinforcement: A Human Cadaveric Weight-Bearing Computed Tomography Study.Bioengineering (Basel, Switzerland) · 2026
- Influence of implant depth on talus-implant biomechanical behavior in partial talus replacement: a finite element analysis.BMC musculoskeletal disorders · 2026
- Development and Biomechanical Study of Micro Anatomic Universal Locking Plate for Talus Fractures.Orthopaedic surgery · 2026
- Motion-Preserving Salvage for Failed Total Ankle Arthroplasty or Talar AVN: Outcomes of Combined Total Ankle Arthroplasty with Custom 3D-Printed Total Talus Replacement.Foot & ankle international · 2026
- Subject-specific kinematic analysis of a total ankle arthroplasty implant.Frontiers in bioengineering and biotechnology · 2026
- Development and biomechanical study of a novel tibial external fixation device in synthetic tibial fracture models.Clinical biomechanics (Bristol, Avon) · 2026