Analysis of the application of computerized optokinetic nystagmus analyzer in vision examination.
Meixia Jiang, Yunyun Sun, Jie Hong, Bowen Zhao, Zhaojun Meng, Nan Ma, Cheng Che, Lanna Zhang,
Beijing Tongren Eye Center, Beijing Tongren Hospital, Capital Medical University; Beijing Ophthalmology & Visual Sciences Key Laboratory, No1. Dongjiaominxiang Street, Dongcheng District, Beijing, China. · Department of Management and Marketing, University of Petroleum Huadong, Qingdao, Shandong, China.
Key points
From the paper's conclusion- The computerized OKN analyzer exhibits favorable reliability and practicability for VA testing in patients aged ≥ 6 years across different refractive errors, with accuracy that appears sufficient for screening and general clinical estimation of VA. However, in very young children (3-4 years) and in eyes with normal VA, agreement with chart‑based VA is limited, and further validation is required.
Abstract
OBJECTIVE: To investigate the practicability, accuracy, and consistency of the computerized optokinetic nystagmus (OKN) analyzer for visual acuity (VA) measurement. METHOD: Cross-sectional consistency evaluation study. A total of 219 subjects (13.48 ± 12.89 years, 115 males, 104 females) underwent bilateral VA testing using a computerized OKN analyzer and a traditional OKN drum. Additionally, 104 subjects (12.58 ± 12.05 years, 54 males, 50 females) were tested bilaterally with the computerized OKN analyzer and a standard logarithmic E chart (random order), including 80 myopic eyes (spherical equivalent [SE] ≤ -0.5 D) and 63 hyperopic eyes (SE ≥ + 0.5 D); VA was recorded in logMAR. Concordance rates between the OKN analyzer and OKN drum were calculated. Intraclass correlation coefficients (ICC) and Bland-Altman plots were used to assess the consistency between the OKN analyzer and a standard logarithmic E chart. Subjects completing the lightbox test were stratified by age (3-4 years, 5 years, 6 years, 7-18 years, 18-58 years), refractive error, and VA range to compare inter-method correlations. RESULTS: Compared with the OKN drum, the computerized OKN analyzer achieved a concordance rate of 74.00% (right eye) and 77.08% (left eye), indicating good agreement. For VA measurements, mean VA (logMAR) of the right eye was 0.24 ± 0.29 (OKN analyzer) vs. 0.24 ± 0.28 (lightbox); left eye was 0.25 ± 0.28 vs. 0.24 ± 0.29. ICC values were 0.92 (right eye) and 0.88 (left eye, both P < 0.001), confirming significant consistency. Bland-Altman analysis showed 95% limits of agreement of -0.228 to 0.233 (right eye) and - 0.266 to 0.293 (left eye). Inter-method consistency improved with age (highest in ≥ 18 years) and was stronger in myopic vs. hyperopic eyes. Notably, correlation weakened as VA improved, with the highest concordance in the 0.1-0.5 decimal VA. CONCLUSION: The computerized OKN analyzer exhibits favorable reliability and practicability for VA testing in patients aged ≥ 6 years across different refractive errors, with accuracy that appears sufficient for screening and general clinical estimation of VA. However, in very young children (3-4 years) and in eyes with normal VA, agreement with chart‑based VA is limited, and further validation is required.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Diagnostic Utility of Accommodative Measurements by Nott Dynamic Retinoscopy in Children with Unilateral Amblyopia.Ophthalmic & physiological optics : the journal of the British College of Ophthalmic Opticians (Optometrists) · 2026
- Automated and Objective Detection of Binocular Vision with a VR-Based Medical Device.Klinische Monatsblatter fur Augenheilkunde · 2026
- Accuracy and precision of visual acuity tests: A systematic review and meta-analysis across testing strategiesmedRxiv · 2026
- Practical Comparison of a Newly Developed Software-Based Visual Acuity Test and Photoscreening in Japanese Preschool Children.Vision (Basel, Switzerland) · 2026
- Machine Learning-Based Prediction of Cycloplegic Refraction Using the Eyerobo Vision Screener: Design-A Cross-Sectional Comparative Device Study.Ophthalmology and therapy · 2026
- Online circular contrast perimetry performed in glaucoma patients in Sub-Saharan Africa: a multi-centre comparative study in Nigeria.Eye (London, England) · 2026