Impact of contact-pressure based Hotballoon ablation of atrial fibrillation.
Shunsuke Uetake, Noriyuki Kobayashi, Nobuaki Ito, Taiji Okada, Toshiki Arai, Rei Mimuro, Nobuaki Kobayashi, Kuniya Asai,
Department of Cardiovascular Medicine, Nippon Medical School Chiba Hokusoh Hospital, 1715 Kamagari, Inzai, Chiba, 270-1694, Japan. s00-015@nms.ac.jp. · Department of Cardiovascular Medicine, Kawaguchi Cardiovascular and Respiratory Hospital, Kawaguchi, Japan.
Key points
From the paper's conclusion- Intra-balloon pressure measurement can be a practical surrogate for balloon-tissue contact and may support a safe, effective, and potentially durable single-shot HotBalloon PVI.
Abstract
BACKGROUND: Effective pulmonary vein isolation (PVI) using HotBalloon ablation requires adequate tissue contact, but no objective index currently exists to assess contact quality. METHODS: We assessed balloon contact pressure (BCP) by measuring the difference in intra-balloon pressure before and after PV contact, using a pressure transducer connected to the balloon catheter shaft. In Phase 1, we retrospectively analyzed BCP in 25 patients (95 PVs) and compared the values between successful and unsuccessful single-shot PVI cases. Receiver operating characteristic analysis determined optimal cutoff values. In Phase 2, ablation was performed only when BCP exceeded the optimal threshold identified in Phase 1. RESULTS: BCP was significantly higher in successful PVI cases for both left superior pulmonary veins (LSPVs) (86.0 ± 27.0 mmHg vs. 24.3 ± 16.1 mmHg, P < 0.001) and non-LSPVs (61.1 ± 22.3 mmHg vs. 24.8 ± 18.6 mmHg, P < 0.001). ROC analysis identified an optimal cutoff value of 40 mmHg for LSPVs and 35 mmHg for non-LSPVs. In Phase 2, the single-shot PVI success rate was 89.3% (25/28 PVs) for LSPVs and 97.6% (82/84 PVs) for non-LSPVs. In addition to sufficient BCP, contact width > 4.2 mm at the carina was also associated with successful PVI. During follow-up, atrial fibrillation recurrence was observed in a subset of patients; however, PV reconnection was infrequently observed among those undergoing repeat ablation. No cases of severe PV stenosis were observed. CONCLUSIONS: Intra-balloon pressure measurement can be a practical surrogate for balloon-tissue contact and may support a safe, effective, and potentially durable single-shot HotBalloon PVI.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Contact Index–Guided Pulsed Field Ablation: A Two‐Phase Validation of the Impedance Change Threshold to Prevent Acute Pulmonary Vein GapsJournal of Arrhythmia · 2026
- Clinical validation of a ≥35% contact index threshold for fixed-loop circular pulsed field ablationEuropean Heart Journal · 2026
- Pressure-Waveform Guidance for Pulmonary Vein Occlusion During Cryoballoon Ablation: Procedural Efficiency and Repeat-Procedure Findings from a Large Single-Center Cohort.Journal of clinical medicine · 2026
- First-in-human experience and one-year outcomes using a novel conformal pulsed-field ablation balloon catheter for pulmonary vein isolation in paroxysmal atrial fibrillation.Journal of interventional cardiac electrophysiology : an international journal of arrhythmias and pacing · 2026
- Development and Validation of a Novel Contact Force-Integrated Pulsed Field Quantization Index for Pulmonary Vein Isolation in a Porcine Model.Journal of cardiovascular electrophysiology · 2026
- Proof of concept - retrospective virtual reality pulmonary veins modeling evaluating eligibility for atrial fibrillation cryoballoon ablationEuropean Heart Journal · 2026