First-in-human experience and one-year outcomes using a novel conformal pulsed-field ablation balloon catheter for pulmonary vein isolation in paroxysmal atrial fibrillation.
Shiqiang Xiong, Duan Luo, Yan Luo, Guoshu Yang, Guijun He, Yongxin Yang, Jialing He, Bao Xu,
Department of Cardiology, Affiliated Hospital of Southwest Jiaotong University, The Third People's Hospital of Chengdu, Chengdu, Sichuan, China. · EnChannel Medical Inc, Guangzhou, China.
Key points
From the paper's conclusion- The conformal PFBalloon catheter demonstrated favorable safety, efficiency, and effectiveness for PVI in PAF patients, achieving durable lesions and a high rate of freedom from atrial arrhythmias recurrence.
Abstract
BACKGROUND: Thermal balloon catheters have improved the efficiency of pulmonary vein isolation (PVI) in atrial fibrillation (AF). Reports assessing additional safety and effectiveness of pulsed field ablation (PFA) balloon catheters for PVI are limited. This study aims to assess the safety and effectiveness of a conformal PFA balloon catheter for PVI in paroxysmal AF. METHODS: 35 patients with paroxysmal AF (PAF) were treated with a PFBalloon catheter (Biphasic, Bipolar, 750 V). The balloon was inflated with 10:1 saline/contrast, with volume adjusted for spherical wide-antral or pear-shaped ostial-junction PV ablation to ensure optimal tissue contact. The primary safety endpoint was absence of periprocedural primary adverse events. The primary effectiveness endpoint was acute PVI. The secondary effectiveness endpoint included durable PVI assessed by remapping at three months post-ablation and 1-year clinical outcomes. RESULTS: All targeted PVs (142/142) in 35 patients were acutely isolated, with 32.3 ± 4.5 applications per patient and 7.9 ± 1.0 per vein. Procedure and left atrial dwell times were 133 ± 46 and 56 ± 25 min, respectively. No hemolysis-related acute kidney injury or severe device-related complications occurred. PVI durability was 76.0% (19/25) per patient and 93.1% (95/102) per vein. After a median follow-up of 367 (interquartile range 357-380) days, 1-year Kaplan-Meier estimate for freedom from atrial arrhythmia (AF/atrial flutter/atrial tachycardia) ≥30 s was 90.7% (95% CI, 81.2%-100%). CONCLUSIONS: The conformal PFBalloon catheter demonstrated favorable safety, efficiency, and effectiveness for PVI in PAF patients, achieving durable lesions and a high rate of freedom from atrial arrhythmias recurrence. TRIAL REGISTRATION: clinicaltrials.gov NCT07320976.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Safety and feasibility of a balloon-based pulsed field ablation system for the catheter ablation of paroxysmal atrial fibrillation.Journal of interventional cardiac electrophysiology : an international journal of arrhythmias and pacing · 2026
- The Pear Strategy for Two-Ring Pulmonary Vein Isolation: Antral and Sleeve Ablation Using a Balloon-in-Basket Pulsed Field Ablation Catheter.Journal of cardiovascular electrophysiology · 2026
- Real-World Experience With the Balloon-in-Basket Pulsed Field Ablation System for Pulmonary Vein Isolation and Ablation of Extra-Pulmonary Vein Sites.Journal of cardiovascular electrophysiology · 2026
- Safety and Effectiveness of a Novel Dual-Energy Radiofrequency/Pulsed Field Ablation Catheter: 12 Month Results of the FlexPulse Study.Europace : European pacing, arrhythmias, and cardiac electrophysiology : journal of the working groups on cardiac pacing, arrhythmias, and cardiac cellular electrophysiology of the European Society of Cardiology · 2026
- Pulsed-Field Ablation to Enhance the Durability of Pulmonary Vein Isolation and Treat Paroxysmal Atrial FibrillationJournal of the American College of Cardiology · 2026
- Pulsed-field ablation demonstrates superior pulmonary vein isolation durability at 90 days: Pooled analysis of protocol-mandated remapping dataHeart Rhythm · 2025