A preliminary assessment of a novel antiviral air filter for continuous end-tidal sampling of carbon dioxide for respiratory monitoring of sedated patients: a pilot study.
Daryl Jian An Tan, Alex Chin Koon Koh, Yew Weng Chan
Division of Anaesthesiology and Perioperative Medicine, Singapore General Hospital, Outram Road, Singapore, 169608, Singapore. · Department of Anaesthesiology, Sengkang General Hospital, 110 Sengkang E Way, Singapore, 544886, Singapore.
Key points
From the paper's conclusion- This pilot feasibility study demonstrates the antiviral filter's potential as an inexpensive, effective, and reliable device for end-tidal CO2 monitoring in clinical setting. Our findings also advocate for its role in promoting sustainable healthcare practices.
Abstract
BACKGROUND: Monitoring expired end-tidal carbon dioxide (CO2) is crucial for sedated patients undergoing various procedures. However, a significant concern is the potential contamination of CO2 sampling lines and gas analyzers when used on patients with respiratory tract infections. To address this, a novel antiviral filter was developed featuring an antimicrobial compound synthesized from cashew skin and husks. Our study aims to assess the device's clinical feasibility for continuous end-tidal CO2 monitoring. METHODS: We conducted a preliminary assessment on 15 adult patients undergoing minor upper limb surgical procedures with local anesthesia, where continuous end-tidal CO2 monitoring is essential. In all cases, the antiviral filter was positioned above the patient's simple face mask and remained there throughout the entire procedure. The primary outcome was the incidence end-tidal CO2 capnogram signal disruptions. The secondary outcome was the occurrence of any patient-reported adverse events during device utilization. RESULTS: The median [range] duration of end-tidal CO2 monitoring using the antiviral filter was 25 [12-38] minutes. Independent observers found no disruptions in the continuous end-tidal CO2 capnograms across all patients. Furthermore, no patients reported any discomfort or foul smell from the antiviral filter. CONCLUSION: This pilot feasibility study demonstrates the antiviral filter's potential as an inexpensive, effective, and reliable device for end-tidal CO2 monitoring in clinical setting. Our findings also advocate for its role in promoting sustainable healthcare practices.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- A Fully Integrated Smart Facemask for Ultrafast and Humidity-Resistant Exhaled CO2 Monitoring.ACS sensors · 2026
- P128. Bacterial viability in PAP airflow after use and reprocessingSLEEP advances · 2026
- Can isolated negative-pressure devices reduce transmission of respiratory infection in critical care units: A laboratory-based controlled experimental observational study.Journal of infection prevention · 2026
- Evaluation of Two Cyclonic Bioaerosol Samplers for Hospital Air Monitoring in a Calibrated Test Chamber Integrating Aerodynamics and Biological EndpointsIndoor Air · 2026
- Personalized automatic management of tracheal cuff pressure and subglottic secretions drainage to prevent pneumonia in critically ill intubated patients. The MICROINHALO multicenter randomized controlled trialIntensive Care Medicine · 2026
- Short-term safety and local tolerability of repeated exposure to aerosolized hypochlorous acid dry fog in healthy adults: a randomized controlled pilot study.Frontiers in public health · 2026