Manual compared to automated closed-loop oxygen control strategies in hospitalised adults: a systematic review and meta-analysis.
Louis Kirton, Rowan Biggs, Neakiry Kivi, Francesca Lynch, Tanira Kingi, Ross Sayers, Jenny Sparks, Allie Eathorne,
The Medical Research Institute of New Zealand, Wellington, New Zealand louis.kirton@mrinz.ac.nz. · The Medical Research Institute of New Zealand, Wellington, New Zealand.
Key points
From the paper's conclusion- CLOC improves adherence to a target S pO2 range compared to manual oxygen control and is an effective approach for achieving a target S pO2 range when delivering oxygen noninvasively in hospitalised adults.
Abstract
BACKGROUND: Closed-loop oxygen control (CLOC) systems improve accuracy of oxygen delivery compared with standard manual oxygen adjustment. Determination of their efficacy in delivering oxygen noninvasively to adults, using different delivery methods and across different hospital locations, would enhance clinical practice. METHODS: We conducted a systematic review and meta-analysis, searching Medline, Embase, Scopus Basic and Advanced, and Cochrane databases. Randomised controlled trials comparing time spent within a target peripheral oxygen saturation (S pO2 ) range between CLOC and manual oxygen control in hospitalised adults (>18 years) were included. Outpatient studies and studies using mechanical ventilation were excluded. The primary outcome was proportion of time spent within a target S pO2 range. Meta-regression was performed for pre-determined study-level covariates, including oxygen delivery mode and locations of care. RESULTS: The search identified 3777 papers; of these, seven were included, with 635 participants. The mean proportion of time spent within the target S pO2 range was between 81.2 and 94.9% for CLOC and 51.3-84.0% for manual oxygen control. The mean difference was 21.2% (95% CI 18.5-23.8, p<0.0001), favouring CLOC. Evidence of heterogeneity (I2=87.9) was explained by three studies using low-flow oxygen therapy, mean difference 30.9% (95% CI 27.0-34.8, p<0.0001); compared with four studies using high-flow nasal oxygen therapy, mean difference 12.7% (95% CI 9.1-16.4, p<0.001). CONCLUSIONS: CLOC improves adherence to a target S pO2 range compared to manual oxygen control and is an effective approach for achieving a target S pO2 range when delivering oxygen noninvasively in hospitalised adults.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Automated Control of Oxygen Therapy: A State-of-the-Art Review of Closed-Loop Devices, with a Rule-Based, Network-Independent System (EMILY) as a Working Exemplar.Seminars in respiratory and critical care medicine · 2026
- Comparison of automated and manual oxygen titration in emergency care: a systematic review and meta-analysis.CJEM · 2026
- Closed-loop insulin delivery for glycaemic control in hospitalised and perioperative adults: A systematic review and meta-analysis of randomised controlled trials.Diabetes research and clinical practice · 2026
- Closed-Loop Insulin Delivery in Inpatient Care: A Systematic Review and Meta-Analysis of Clinical and Implementation Outcomes.Diabetes, obesity & metabolism · 2026
- Comparison of two automated oxygen control algorithms in mechanically ventilated neonates: an open label, randomised, crossover trial.Archives of disease in childhood. Fetal and neonatal edition · 2026
- AI-Driven and Automated Systems for Continuous Oxygen Saturation Monitoring in Long-Term Oxygen Therapy: Systematic Review.JMIRx med · 2026