◆ Archives of disease in childhood. Fetal and neonatal edition2026-08-27

Comparison of two automated oxygen control algorithms in mechanically ventilated neonates: an open label, randomised, crossover trial.

Sumit Bhadu, Venkataseshan Sundaram, Jogender Kumar, Praveen Kumar

一句话结论 · In one sentence

In mechanically ventilated neonates, adaptive CLiO2 algorithm did maintain the oxygen saturation within target range for a greater proportion of time and reduced hyperoxic exposure compared with the rule-based PRICO algorithm.

原始摘要(英文原文)· Original abstract
OBJECTIVE: To compare rule-based non-fuzzy algorithm (Predictive Intelligent Control of Oxygenation (PRICO)) with adaptive algorithm (Closed-loop Inspired Oxygen (CLiO2)) for automated oxygen control in mechanically ventilated neonates. DESIGN: Open label, randomised, crossover trial. SETTING: Level III neonatal intensive care unit. PATIENTS: Neonates ≤44 weeks post-menstrual age receiving invasive or non-invasive ventilation and fraction of inspired oxygen (FiO₂) ≥25%. INTERVENTIONS: Participants randomised to one of two sequences of auto-FiO2 (CLiO2 then PRICO or PRICO then CLiO2). Each algorithm in the sequence was applied for 24 hours before crossing over to an alternate algorithm. Target saturation was 91-95%. Masimo pulse-oximeters were used for saturation measurement. MAIN OUTCOME: Primary outcome was time-weighted proportion of time spent within target saturation range. Secondary outcomes were time spent below and above target saturation range. RESULTS: 32 neonates were enrolled; 29 completed both intervention periods and contributed to paired data. Mean time-weighted proportion within target range was 76% in the CLiO2 group compared with 62% in PRICO (mean within-participant difference (PRICO-CLiO2) -14 percentage points, 95% CI -24 to -4; paired t-test p=0.01). A mixed-effects model accounting for period gave an 18 percentage points difference (82% vs 64%; treatment effect -0.92 on the logit scale, 95% CI -1.51 to -0.33, p=0.004). Time-weighted proportion spent above target range was lower with CLiO2 (15% vs 30%). Time spent below target range was comparable between groups. No significant period or carryover effect was observed. CONCLUSIONS: In mechanically ventilated neonates, adaptive CLiO2 algorithm did maintain the oxygen saturation within target range for a greater proportion of time and reduced hyperoxic exposure compared with the rule-based PRICO algorithm. TRIAL REGISTRATION NUMBER: CTRI/2025/05/087301.
读原文 · Read the paper ↗

AI 追问PRO

登录后使用 AI 追问

讨论区

登录后参与讨论

相关论文 · Related

Comparison of two automated oxygen control algorithms in mechanically ventilated neonates: an open label, randomised, crossover trial. — 科研速览 Science Skim