Comparison of two automated oxygen control algorithms in mechanically ventilated neonates: an open label, randomised, crossover trial.
Sumit Bhadu, Venkataseshan Sundaram, Jogender Kumar, Praveen Kumar
Pediatrics, Post Graduate Institute of Medical Education and Research, Chandigarh, Chandigarh, India. · Division of Neonatology, Pediatrics, Post Graduate Institute of Medical Education and Research, Chandigarh, Chandigarh, India venkatpgi@gmail.com.
Key points
From the paper's conclusion- In mechanically ventilated neonates, adaptive CLiO2 algorithm did maintain the oxygen saturation within target range for a greater proportion of time and reduced hyperoxic exposure compared with the rule-based PRICO algorithm.
Abstract
OBJECTIVE: To compare rule-based non-fuzzy algorithm (Predictive Intelligent Control of Oxygenation (PRICO)) with adaptive algorithm (Closed-loop Inspired Oxygen (CLiO2)) for automated oxygen control in mechanically ventilated neonates. DESIGN: Open label, randomised, crossover trial. SETTING: Level III neonatal intensive care unit. PATIENTS: Neonates ≤44 weeks post-menstrual age receiving invasive or non-invasive ventilation and fraction of inspired oxygen (FiO₂) ≥25%. INTERVENTIONS: Participants randomised to one of two sequences of auto-FiO2 (CLiO2 then PRICO or PRICO then CLiO2). Each algorithm in the sequence was applied for 24 hours before crossing over to an alternate algorithm. Target saturation was 91-95%. Masimo pulse-oximeters were used for saturation measurement. MAIN OUTCOME: Primary outcome was time-weighted proportion of time spent within target saturation range. Secondary outcomes were time spent below and above target saturation range. RESULTS: 32 neonates were enrolled; 29 completed both intervention periods and contributed to paired data. Mean time-weighted proportion within target range was 76% in the CLiO2 group compared with 62% in PRICO (mean within-participant difference (PRICO-CLiO2) -14 percentage points, 95% CI -24 to -4; paired t-test p=0.01). A mixed-effects model accounting for period gave an 18 percentage points difference (82% vs 64%; treatment effect -0.92 on the logit scale, 95% CI -1.51 to -0.33, p=0.004). Time-weighted proportion spent above target range was lower with CLiO2 (15% vs 30%). Time spent below target range was comparable between groups. No significant period or carryover effect was observed. CONCLUSIONS: In mechanically ventilated neonates, adaptive CLiO2 algorithm did maintain the oxygen saturation within target range for a greater proportion of time and reduced hyperoxic exposure compared with the rule-based PRICO algorithm. TRIAL REGISTRATION NUMBER: CTRI/2025/05/087301.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Neonatal automated oxygen titration algorithms: response characteristics and performance.Archives of disease in childhood. Fetal and neonatal edition · 2026
- Automatic versus manual control of oxygen and neonatal clinical outcomes in extremely preterm infants: a multicentre, parallel-group, randomised, controlled, superiority trialThe Lancet Child & Adolescent Health · 2026
- Effect of small changes in set oxygen saturation target range during routine use of automated FiO2 control: a randomized physiological study.Frontiers in pediatrics · 2026
- Automated Control of Oxygen Therapy: A State-of-the-Art Review of Closed-Loop Devices, with a Rule-Based, Network-Independent System (EMILY) as a Working Exemplar.Seminars in respiratory and critical care medicine · 2026
- Manual compared to automated closed-loop oxygen control strategies in hospitalised adults: a systematic review and meta-analysis.European respiratory review : an official journal of the European Respiratory Society · 2026
- Oxygenation Indices with Oxygen Challenge Test in Neonates During VA ECMO.Children (Basel, Switzerland) · 2026