Swinging vs Flicking: Assessing Safer Removal Techniques for Air Bubbles in Intravitreal Injections.
Paula Sakemi Fukuhara, Alexandra Lucas, Krishna M G Mallela, Øystein Kalsnes Jørstad, Michel Eid Farah, Gustavo Barreto Melo
Department of Ophthalmology, Federal University of São Paulo, São Paulo, Brazil. · Center for Pharmaceutical Biotechnology, Skaggs School of Pharmacy and Pharmaceutical Sciences, University of Colorado Anschutz Medical Campus, Aurora, CO, USA.
Key points
From the paper's conclusion- Compared with flicking, a gentle swinging motion reduced particle release across syringe models while allowing air bubbles to be removed. This technique may represent a simple modification to intravitreal injection preparation that could reduce particulate contamination.
Abstract
PURPOSE: To evaluate whether a gentler swinging motion, rather than flicking, can effectively eliminate air bubbles from syringes before intravitreal injections while minimizing particle generation. METHODS: Three syringe models were assessed under controlled conditions: Zero Residual Silicone Oil (SO)-free 0.2 mL syringe, BD Tuberculin 1 mL slip-tip syringe (siliconized syringe 1), and HSW Soft-Ject 1 mL Luer-Lock syringe (siliconized syringe 2). Two solutions were tested: a polysorbate-based buffer and bevacizumab (Avastin). Each syringe was subjected to 3 handling conditions: no agitation (control), swinging (hand-induced centrifugal movement), and flicking. The number of released particles was quantified using FlowCam imaging, and statistical analyses were conducted to compare handling techniques and syringe models. RESULTS: Flicking significantly increased particle release compared with both swinging and the control condition (P < .01). Swinging consistently resulted in lower particle counts across all syringe models compared with flicking, although the decrease in magnitude varied. The SO-free syringe demonstrated the lowest overall particle release, whereas siliconized syringes exhibited greater variability in particle release. Particle distribution also differed between the buffer and bevacizumab solutions, suggesting that fluid composition influences particle formation and release. CONCLUSIONS: Compared with flicking, a gentle swinging motion reduced particle release across syringe models while allowing air bubbles to be removed. This technique may represent a simple modification to intravitreal injection preparation that could reduce particulate contamination.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Performance characterization of syringes used for intravitreal injections: dosing accuracy, particulate contamination, and residual volume.International journal of retina and vitreous · 2026
- Physical Parameters Influencing the Rate of Subretinal Injection: A Microfluidic Analysis.Translational vision science & technology · 2026
- Mechanical stresses on the retina induced by intravitreal injections in an artificial model of the vitreous chamberGraefe s Archive for Clinical and Experimental Ophthalmology · 2026
- Effects of Siliconization Techniques, Headspace, Mechanical and Thermal Stress on Particle Generation in an Antibody Formulation.Pharmaceutical research · 2026
- Preclinical Evaluation of a Novel Method and Device for Performing Minimally Invasive Suprachoroidal Hydrogel Buckling.Clinical & experimental ophthalmology · 2026
- Factors affecting the longevity and quality of detergent sclerosant foams.Phlebology · 2026