Clinical & experimental ophthalmology

Preclinical Evaluation of a Novel Method and Device for Performing Minimally Invasive Suprachoroidal Hydrogel Buckling.

Gareth D Mercer, Yoreh Barak, Keren Mano Tamir, Peter J Kertes, Kenneth T Eng

Department of Ophthalmology and Vision Sciences, Temerty Faculty of Medicine, University of Toronto, Toronto, Canada. · Department of Ophthalmology, Rambam Health Care Campus, Haifa, Israel.

Key points

From the paper's conclusion
  • Conclusions This study establishes the technical feasibility of using a novel injector to perform minimally invasive SCS hydrogel buckling and provides estimates of expected buckle dimensions across a range of clinically relevant injection volumes.

Abstract

BACKGROUND: We evaluated a novel device and technique for performing suprachoroidal buckling and modelled the relationship between injection volume and buckle morphology. The device utilises a non-sharp tissue separator which, via tangential blunt dissection, opens a channel to the suprachoroidal space (SCS). METHODS: We injected sodium hyaluronate 2.3% into the SCS of fresh whole human eyes (n = 34), evaluating injection volumes of 0.1, 0.2 and 0.3 mL and needle insertion positions of 6, 10 and 15 mm posterior to the limbus, and measured the resulting buckle dimensions using B-scan ultrasonography. RESULTS: All but one injection resulted in successful SCS buckle creation. We observed a linear relationship between injection volume and buckle height, transverse basal diameter (TBD) and longitudinal basal diameter (LBD; R2 0.64, 0.65 and 0.64). A 0.1 mL injection resulted in a buckle with average height of 2.5 mm (95% confidence interval, CI: 2.3-2.8), TBD of 9.0 mm (95% CI: 8.5-9.5) and LBD of 8.6 mm (95% CI: 8.1-9.2). A 0.1 mL increase in volume was associated with an average height increase of 0.8 mm (95% CI: 0.6-1.0), TBD increase of 1.8 mm (95% CI: 1.3-2.2) and LBD increase of 1.6 mm (95% CI: 1.2-2.1). Needle insertion distance did not significantly modify these relationships. CONCLUSIONS: This study establishes the technical feasibility of using a novel injector to perform minimally invasive SCS hydrogel buckling and provides estimates of expected buckle dimensions across a range of clinically relevant injection volumes.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers

Preclinical Evaluation of a Novel Method and Device for Performing Minimally Invasive Suprachoroidal Hydrogel Buckling — Science Skim