Journal of vitreoretinal diseases

Visualization Techniques for Port Delivery System Refill-Exchange.

Mitchell T Allphin, Carl C Awh

Tennessee Retina, Nashville, TN, USA.

Key points

From the paper's conclusion
  • Indirect internal illumination of the PDS implant can facilitate visualization of the implant septum during the refill-exchange procedure, particularly when visualization is compromised by conjunctival edema, thickened Tenon capsule, or fibrous capsule formation.

Abstract

PURPOSE: To describe methods for improving visualization of the port delivery system (PDS) with ranibizumab (Susvimo) during the refill-exchange procedure when visualization of the implant septum is compromised. METHODS: Multiple methods of PDS visualization using external illumination were evaluated. RESULTS: Transpupillary and transscleral illumination of the PDS reservoir body using various light sources improved visualization of the implant septum during the refill-exchange procedure. CONCLUSIONS: Indirect internal illumination of the PDS implant can facilitate visualization of the implant septum during the refill-exchange procedure, particularly when visualization is compromised by conjunctival edema, thickened Tenon capsule, or fibrous capsule formation.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers