End-Tidal Propofol Monitor Based on Deep Ultraviolet Fluorescence Spectroscopy Using Physics-Based Real-Time Correction and Deconvolution Models

Akio Shimono, Toshihide Hikida, Asao Mizoguchi, Takahiro Goto, Kenichi Masui

Key points

From the paper's conclusion
  • We successfully demonstrated breath-by-breath detection sensitivity down to 0.3 ppb for end-tidal propofol concentration with minimal calibration, indicating that a non-invasive end-tidal propofol monitor as a medical device is feasible.

Abstract

Since the 2007 report of a correlation between exhaled and plasma propofol using proton transfer reaction mass spectrometry, there has been a desire to develop clinically acceptable exhaled propofol monitors that are compact, simple, practical, and economical. Since the propofol concentration in the breathing circuit during anesthesia fluctuates significantly every 5−10 s, monitors used in anesthesia practice require high time-resolution and highly sensitive measurements to enable continuous monitoring. Additionally, the composition of the gas sampled from the breathing circuit changes frequently due to bidirectional flow in the circuit, drastic changes in matrix composition between inhaled and exhaled airs, especially water vapor, and the adsorption and desorption of propofol on the inner surfaces. A propofol concentration waveform is also required to confirm measurement validity in clinical practice, similar to a capnometer that continuously displays high-resolution waveforms. In this study, we developed a fluorescence gas instrument with a time resolution of 10 Hz using deep ultraviolet LED excitation. We implemented a real-time algorithm to precisely calculate and correct background signal and fluorescence quenching using physical models, while monitoring major matrix components with auxiliary sensors. The end-tidal propofol concentration of every ventilation period was then determined by applying a first-order lag model to propofol in the sidestream sampling. We successfully demonstrated breath-by-breath detection sensitivity down to 0.3 ppb for end-tidal propofol concentration with minimal calibration, indicating that a non-invasive end-tidal propofol monitor as a medical device is feasible.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers