Phantom-Based Study on Testicular Dose Protection Using Different Thicknesses of Low-Melting-Point Lead Shields in Postoperative Radiotherapy for Seminoma.
Li Zou, Huanhuan Fei, Wenlang Zhu, Haoyu Wang, Gang Zhou, Ye Tian
Department of Radiotherapy & Oncology, The Second Affiliated Hospital of Soochow University, Suzhou, China. · Department of Gastroenterology, Nanjing First Hospital, Nanjing Medical University, Nanjing, China.
Key points
From the paper's conclusion- Customized low-melting-point lead shields can effectively reduce contralateral testicular scattered dose in a phantom-based seminoma radiotherapy setup. The dose reduction depends on both shield thickness and distance from the radiation field edge.
Abstract
BACKGROUND: This phantom-based study evaluated the effects of low-melting-point lead shield thickness and field-edge distance on contralateral testicular dose. METHODS: A 3D-CRT plan was generated de novo on CT images of the SZMC-IV anthropomorphic phantom; no patient-derived imaging or treatment-plan data were used. A 0.6 cm3 thimble ionization chamber measured dose per fraction at different field-edge distances without shielding and with 3, 5, 7, 10, and 15 mm lead shields. Measurement-derived normalized doses were fitted using an exponential model. The 36 Gy scenario was evaluated only by linear scaling. RESULTS: Normalized testicular dose decreased exponentially with increasing field-edge distance and shield thickness. For the nominal 26 Gy scenario, measurement-derived cumulative doses were 0.315 Gy with 3 mm shielding at 10 cm and 0.481 Gy with 7 mm shielding at 4 cm. For the hypothetical 36 Gy scenario, the measurement-based linearly scaled estimate was 0.432 Gy with 15 mm shielding at 4 cm. Excluding non-independent 36 Gy estimates, the RMSE was 0.074 Gy. CONCLUSIONS: Customized low-melting-point lead shields can effectively reduce contralateral testicular scattered dose in a phantom-based seminoma radiotherapy setup. The dose reduction depends on both shield thickness and distance from the radiation field edge.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Evaluation of testicular shielding effectiveness for surface and internal photon dose using a 3D-printed scrotal phantom.Journal of applied clinical medical physics · 2026
- Long‐term complete remission following low‐dose (20 Gy) radiation therapy for extensive metastatic seminoma in a dogVeterinary Record Case Reports · 2026
- External dose characteristics and radiation-protection metrics after permanent 125I prostate brachytherapy: comparison of conventional and sparse implantation techniques.Radiation protection dosimetry · 2026
- Evaluation of wedge-induced photoneutron dose during 15 MV 3D conformal rectal radiotherapy using TLD600/TLD700 pairs.Journal of X-ray science and technology · 2026
- Effects of varying lateral spot size and beam energy spread in proton therapy: a phantom and patient case study.Physics in medicine and biology · 2026
- Dosimetric impact of computational phantom selection in passive scattering proton therapy: Development and validation of an extended MIRD prostate phantom using Geant4 Monte Carlo simulation.Radiation and environmental biophysics · 2026