Dosimetric impact of computational phantom selection in passive scattering proton therapy: Development and validation of an extended MIRD prostate phantom using Geant4 Monte Carlo simulation.
Mostafa Khosravi, Mohammad Torkiha, Daryoush Shahbazi-Gahrourei
Department of Physics, Kashan University, Kashan, Iran. · Department of Medical Physics, School of Medicine, Isfahan University of Medical Sciences, Isfahan, 81746-73461, Iran. shahbazi@med.mui.ac.ir.
Key points
From the paper's conclusion- These findings suggest that computational phantom selection represents an important modelling factor influencing Monte Carlo proton dose prediction and that mathematically defined phantoms can provide a flexible platform for methodological and dosimetric investigations when their inherent anatomical simplifications are appropriately considered.
Abstract
Accurate anatomical modelling is important for reliable Monte Carlo (MC) proton dosimetry, yet the influence of computational phantom selection on clinically relevant dose prediction remains insufficiently quantified. In this study, an anatomically extended Medical Internal Radiation Dose (MIRD) computational phantom was developed for passive scattering proton therapy of prostate cancer and systematically compared with the anatomically realistic reference male phantom recommended by the International Commission on Radiological Protection using a validated Geant4 MC simulation framework. The extended phantom incorporated the prostate, bladder, rectum, urethra, and intraprostatic tumour volumes. At the same time, a clinical passive-scattering beamline at the CATANA proton therapy facility was modelled to generate spread-out Bragg peaks for treatment planning. The simulation framework demonstrated excellent agreement with experimental depth-dose measurements, yielding a practical proton range (R90) deviation of 1.10 mm and a gamma passing rate of 98.63% under the 3%/2 mm criterion, confirming the accuracy of the dosimetric model. Following dose normalisation, both phantom models achieved identical target coverage (D95=74 Gy) and conformity (CI = 0.95). In contrast, anatomical representation produced measurable differences in proton transport characteristics, including a 5.29% variation in proton range and a 72.4% difference in distal falloff. Quantitative dosimetric analysis further demonstrated differences in internal target dose distribution and organ-at-risk dose metrics despite comparable overall treatment quality, indicating that conventional target coverage indices alone do not fully reflect the dosimetric consequences of anatomical modelling. Although all evaluated organ-at-risk doses remained below commonly accepted clinical dose constraints, the observed variations in dose deposition highlight the sensitivity of proton transport to tissue geometry and heterogeneity. These findings suggest that computational phantom selection represents an important modelling factor influencing Monte Carlo proton dose prediction and that mathematically defined phantoms can provide a flexible platform for methodological and dosimetric investigations when their inherent anatomical simplifications are appropriately considered.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Validation of a neonatal computational anthropomorphic phantom using Monte Carlo simulation.Radiation protection dosimetry · 2026
- Dosimetric evaluation and energy optimization in proton therapy for Hodgkin’s lymphoma using a GATE-based Monte Carlo frameworkScientific Reports · 2026
- Automated system for MCNP input generation and ICRP 110 phantom adaptation in 125I prostate brachytherapy with edema.Applied radiation and isotopes : including data, instrumentation and methods for use in agriculture, industry and medicine · 2026
- Electron specific absorbed fractions for the ICRP family of mesh-type reference computational phantoms.Physics in medicine and biology · 2026
- Dose distribution in a cervix phantom quality-assurance using two computer modeling approaches.Applied radiation and isotopes : including data, instrumentation and methods for use in agriculture, industry and medicine · 2026
- Enhancements and reference internal dosimetry data for the MOBY whole-body computational phantom of the laboratory mouse.Physics in medicine and biology · 2026