Loose Screw or No Screw? A Biomechanical Study to Answer a Common Intraoperative Dilemma.
Christopher C Y Chung, Adina Harri, Katherine Shehadeh, Calvin K Chan, Nicole S Pham, Jeffrey Yao
Stanford University, CA, USA.
Key points
From the paper's conclusion- Removing a single loose screw in bridge plate fixation for comminuted metacarpal diaphyseal fractures safely avoids screw migration without imparting a stress riser or compromising biomechanical strength.
Abstract
BACKGROUND: Insertion of cortical screws with poor purchase during metacarpal plate fixation may occur, yet the biomechanical tradeoff of retaining versus removing a loose screw remains difficult to quantify. Removing a loose screw avoids risking screw migration but may decrease fixation strength or create a stress riser at the empty hole. Our study evaluated bending and torsional properties in plate fixation for comminuted metacarpal diaphyseal fractures with or without a single loose screw. METHODS: Sixty-four cadaveric fractures were fixed in bridge mode using nonlocking screws. A single screw was overtightened and stripped to simulate screw loosening and then either retained or removed. Testing groups were assigned based on loading configuration (cantilever bending or torsion) and the presence of a loose screw. Multivariable linear regression models evaluated relationships between screw presence and biomechanical outcome measures. Failure modes were analyzed using Fisher exact test. RESULTS: Cantilever bending load at 2 mm displacement was 7.6 N (SD = 1.9 N) with loose screw retained versus 8.0 N (SD = 2.6 N) with screw removed (P = .796). Torque at 10° rotational displacement showed no difference between retained (0.363 N m, SD = 0.054 N m) and removed screw groups (0.370 N m, SD = 0.139 N m; P = .858). No differences were identified in stress or stiffness. Cantilever bending clinical failure occurred with all specimens bending through the plate. During torsional testing, 6 of 16 with retained loose screw secondarily failed, with the loose screw hole fracturing after rotating past the clinical endpoint. CONCLUSION: Removing a single loose screw in bridge plate fixation for comminuted metacarpal diaphyseal fractures safely avoids screw migration without imparting a stress riser or compromising biomechanical strength.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Biomechanical comparison of mini-plate and screw fixation in minimally invasive distal metatarsal osteotomy for hallux valgus : a finite element and cadaveric study.Bone & joint research · 2026
- Biomechanical Evaluation of Crossed Screw Fixations With Different Head Designs for First Tarsometatarsal Joint Arthrodesis.Foot & ankle specialist · 2026
- Index Metacarpal Fracture Risk in Suture Button Suspensionplasty and the Effects of Drill Hole Position: A Cadaveric Study.The Journal of hand surgery · 2026
- Biomechanical Comparison of Composite versus Screw-Rod Fixation for Posterior C5-C6 Fusion.Global spine journal · 2026
- Radiological outcomes of metacarpal fractures fixed with intramedullary headless compression screws compared to dorsal plates.Hand surgery & rehabilitation · 2026
- Biomechanical effects of far-cortex over-penetration in locking plate fixation: A finite element study.Proceedings of the Institution of Mechanical Engineers. Part H, Journal of engineering in medicine · 2026