Efficacy and Safety of the MiniMed™ Vivera™ Algorithm: Results from a First-in-Human Feasibility Trial.
Amir Tirosh, Benjamin J Wheeler, Maya Laron Hirsh, Ella Idan, Yael Cohen, Alisa Boucsein, Wipa Warunyuwong, Natalie Kurtz,
Sheba Medical Center, Tel Aviv, Israel. · University of Otago, Otago, New Zealand.
Key points
From the paper's conclusion- In conclusion, this feasibility trial demonstrated that the Vivera algorithm was safe and efficacious in adults with type 1 diabetes, supporting further evaluation in a pivotal trial.
Abstract
MiniMed™ Vivera™ is an automated insulin delivery algorithm designed to make meal announcements optional. This first-in-human feasibility trial evaluated the efficacy and safety of the algorithm under different meal-announcement strategies. Twenty-four adults with type 1 diabetes participated in this prospective study comprising three consecutive 3-week phases: full meal announcement (FA) phase, no meal announcement (NA) phase, and announcement-at-will phase where participants could decide whether to announce each meal (AW). Continuous glucose monitoring data and safety data were assessed for each phase. The international glycemic targets regarding time in, above, and below range were achieved on average across all three meal-announcement phases. Average time in range (70-180 mg/dL) was 82.3 ± 6.60%, 73.8 ± 7.67%, and 76.7 ± 8.98% during FA, NA, and AW phase, respectively. No device-related serious adverse events were reported. In conclusion, this feasibility trial demonstrated that the Vivera algorithm was safe and efficacious in adults with type 1 diabetes, supporting further evaluation in a pivotal trial.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Advanced automated insulin delivery in inpatients receiving nutritional support: a single-center, open-label, randomised controlled trialDiabetes & Metabolism · 2026
- Real-world Glycemic Outcomes of a Tubeless Automated Insulin Delivery System (Medtrum A8 TouchCare) in Adults With Type 1 DiabetesCanadian Journal of Diabetes · 2026
- Impact of DBLG1 Hybrid Closed-Loop Software Optimization and Meal Declaration Pattern on Glucose Metrics in Type 1 Diabetes: A 12-Month Prospective Evaluation in 2014 Users.Diabetes, obesity & metabolism · 2026
- Intrapartum and Early Postpartum Use of Automated Insulin Delivery in Type 1 Diabetes: A Prespecified Analysis of the CIRCUIT Randomized Controlled TrialDiabetes Care · 2026
- The Smart MDI System in People With Type 1 Diabetes and High HBA1C: A Randomized Controlled Trial Demonstrating That Response to Alerts Drives Outcomes.Diabetes, obesity & metabolism · 2026
- MiniMed 780G Advanced Hybrid Closed-Loop System Use During Days Without User-Initiated BolusesDiabetes Care · 2026