Impact of DBLG1 Hybrid Closed-Loop Software Optimization and Meal Declaration Pattern on Glucose Metrics in Type 1 Diabetes: A 12-Month Prospective Evaluation in 2014 Users.
Pierre-Yves Benhamou, Aurélien Vesin, Lucile Reynaud, Thibault Le Roux-Mallouf, Benjamin Chatel
Department of Endocrinology, Grenoble University Hospital, Grenoble Alpes University, INSERM U1055, LBFA, Grenoble, France. · Diabeloop SA, Grenoble, France.
Key points
From the paper's conclusion- This study demonstrates the sustained safety and effectiveness of the latest version of the DBLG1 software, with TIR exceeding 70% throughout the 12-month follow-up period in a large group of 2014 users. Furthermore, while glycaemic control was negatively affected when meal declaration was reduced, the software maintained its safety and acceptable effectiveness in this latter situation.
Abstract
BACKGROUND: The meal period is critical for the outcome of automated insulin delivery. Our aim was to explore the long-term impact on glucose metrics of algorithmic modifications of the DBLG1 hybrid closed-loop system, with special emphasis on the meal declaration pattern. METHODS: We prospectively analysed CGM data from consenting adult patients with Type 1 diabetes across Europe who initiated the new version of DBLG1 paired with the Kaleido patch pump upon its release. Inclusion occurred over 7 months, with follow-up extended up to 12 months. Time in range 70-180 mg/dL (TIR) was the primary endpoint. Utility endpoints included the number of declared meals. RESULTS: This study included 2014 users, age 42.5 ± 13.8 years, baseline HbA1c 7.6% ± 1.2%. Median TIR was 72.0%, 72.3%, 72.0% and 71.7% at M3, M6, M9 and M12, respectively. At M3, time < 54 mg/dL (0.2%), < 70 mg/dL (1.2%) and > 250 mg/dL (5.4%) were controlled and remained stable throughout M12, as well as GMI (7.0%-7.1%). The incidence rates of severe hypo- or hyperglycaemia remained very low. TIR in patients declaring ≥ 3 meals/day exceeded TIR in those declaring < 1 meal/day (73.3 [67.3, 78.7]% vs. 65.7 [60.7, 72.2]%, p < 0.0001). CONCLUSION: This study demonstrates the sustained safety and effectiveness of the latest version of the DBLG1 software, with TIR exceeding 70% throughout the 12-month follow-up period in a large group of 2014 users. Furthermore, while glycaemic control was negatively affected when meal declaration was reduced, the software maintained its safety and acceptable effectiveness in this latter situation.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Prospective multicenter evaluation of glycemic and patient-reported outcomes following transition to a next-generation continuous glucose monitoring system in users of advanced hybrid closed-loop technologyTherapeutic Advances in Endocrinology and Metabolism · 2026
- Development and Validation of a Meal Detection-Based Tool for User Behavior and Meal Management Analysis.Journal of diabetes science and technology · 2026
- Dose-Normalised Time in Range (dn-TIR) Improves Following Hybrid Closed-Loop (HCL) Therapy in Type 1 Diabetes: A Real-World Cohort Study From Australia and the United Kingdom.Diabetes, obesity & metabolism · 2026
- A retrospective observational study evaluating the safety and effectiveness of commercially available automated insulin delivery systems in people with type 1 diabetes and gastroparesisDiabetes Obesity and Metabolism · 2026
- Improvement in Tight Glycemic Control in Adults with Type 1 Diabetes Using the MiniMed™ 780G Hybrid Closed-Loop System: A Real-World Evaluation.Current diabetes reviews · 2026
- MiniMed 780G Advanced Hybrid Closed-Loop System Use During Days Without User-Initiated BolusesDiabetes Care · 2026