Explantation rates and failure mechanisms in early cannulation grafts: A comparative study between Flixene and Gore Acuseal.
Andrea Gattuso, Marco Tadiello, Anna Maria Pisacreta, Silvia Stegher, Marco Heidempergher, Maria Antonietta Orani, Tiziano Porretta
Vascular Surgery Unit, ASST Fatebenefratelli-Sacco, Milan, Italy. · Department of Nephrology, Dialysis and Renal Transplantation, Fondazione IRCCS Cà Granda Ospedale Maggiore Policlinico, Milan, Italy.
Key points
From the paper's conclusion- Gore Acuseal grafts are associated with higher complication and explantation rates compared to Flixene. When analyzed with appropriate survival methodology, time to failure also differs significantly between devices. Distinct patterns of structural complications suggest that graft material and design influence both durability and mechanisms of failure. Further prospective studies are needed to confirm these findings and guide graft selection.
Abstract
BACKGROUND: Early cannulation arteriovenous grafts (ecAVGs) are increasingly used to reduce catheter dependence in hemodialysis patients, but device-specific complications remain a concern. This study aimed to compare explantation rates, timing, and mechanisms of failure between Flixene and Gore Acuseal grafts. METHODS: We conducted a retrospective observational study including patients undergoing prosthetic vascular access implantation. Patients were stratified according to graft type (Flixene vs Gore Acuseal). The primary outcome was graft explantation rate. Secondary outcomes included time to explantation, overall complication rate, and causes of failure. Statistical analyses included chi-square/Fisher's exact tests, Kaplan-Meier survival analysis with log-rank testing incorporating full follow-up data with appropriate censoring, and multivariable logistic regression for infection-related explantation. RESULTS: A total of 69 grafts were analyzed (30 Flixene, 39 Gore Acuseal). Gore Acuseal grafts showed a significantly higher explantation rate (48.7% vs 16.7%, p = 0.006). Kaplan-Meier survival analysis incorporating individual follow-up durations with appropriate censoring revealed significantly different graft survival curves between groups (log-rank p = 0.039). Median graft survival was 39 months for Gore Acuseal; median survival was not reached for Flixene within the observation period. Structural failure was the predominant cause of explantation in both groups. Complication patterns differed significantly: delamination was more frequent in Gore Acuseal grafts (30.8% vs 3.3%), whereas pseudoaneurysm was relatively more common in Flixene grafts (13.3% vs 10.3%; p = 0.015). Infection-related explantation occurred only in the Gore Acuseal group but was not statistically significant. CONCLUSIONS: Gore Acuseal grafts are associated with higher complication and explantation rates compared to Flixene. When analyzed with appropriate survival methodology, time to failure also differs significantly between devices. Distinct patterns of structural complications suggest that graft material and design influence both durability and mechanisms of failure. Further prospective studies are needed to confirm these findings and guide graft selection.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Evaluation and comparison of patency rate and complications of synthetic arteriovenous grafts in the groin and upper extremities among hemodialysis patients.Journal of research in medical sciences : the official journal of Isfahan University of Medical Sciences · 2026
- Completion angiography and venous outflow assessment in arteriovenous graft placement: A retrospective observational study.The journal of vascular access · 2026
- Comparative outcomes of arteriovenous accesses with and without tunneled hemodialysis catheters: A retrospective cohort study.Medicine · 2026
- Real-world outcomes of endoanchors in the prevention and treatment of proximal endograft failure in a multicenter registry.Journal of vascular surgery · 2026
- Dacron vascular grafts: a long-term solution for short renal vein challenges in living donor kidney transplantation in a retrospective case series.Annals of surgical treatment and research · 2026
- Association between early vein graft failure and long-term major adverse cardiovascular events following off-pump coronary artery bypass grafting: an exploratory retrospective cohort study.Journal of thoracic disease · 2026