Real-world outcomes of endoanchors in the prevention and treatment of proximal endograft failure in a multicenter registry.
Michele Antonello, Simona Marrocco, Giovanni Pratesi, Umberto M Bracale, Raffaele Pulli, Diego Moniaci, Stefano Bonvini, Gianfranco Veraldi,
Vascular and Endovascular Surgery Division, Department of Cardiac, Thoracic, Vascular Sciences and Public Health, Padua University, Padua, Italy. · Vascular and Endovascular Surgery Unit, Ospedale Policlinico San Martino, University of Genoa, Genoa, Italy.
Key points
From the paper's conclusion- In this real-world registry, ESAR was used as an adjunct during primary EVAR in patients with a hostile proximal neck and as a salvage treatment of proximal endograft failure after a previous EVAR, with favorable success rates in both settings. In the preventive setting, outcomes were less favorable when the anatomical proximal neck length was <7 mm. In the treatment of type Ia endoleak, favorable results were obtained in up to 80% of cases, with a higher success rate when endoanchors were combined with a proximal aortic extension, and the persistence of a type Ia endoleak on the completion angiogram was a negative prognostic factor. Future comparative studies are necessary to better clarify the clinical efficacy of ESAR in a prevention and treatment setting.
Abstract
OBJECTIVE: To investigate the outcomes of the Heli-FX EndoAnchor system (ESAR) in the prevention and treatment of proximal endograft failure after endovascular aneurysm repair (EVAR). METHODS: We conducted a retrospective data analysis from a multicenter registry of patients treated by ESAR (2017-2025). In the Prevention group (PG), endoanchors were used to prevent proximal endograft failure during primary endovascular repair (EVAR). In the treatment group (TG), endoanchors were used to treat a type Ia endoleak and/or migration after a prior EVAR. Endpoints were technical success, procedural success, and freedom from proximal endograft failure (new-onset type Ia endoleak, related reintervention, endograft migration). RESULTS: There were 325 patients, 234 (72%) in the PG and 91 (28%) in the TG. Technical success was 100% in the PG and 97% in TG (P=.043), with a procedural success of 100% vs 88% (P<.001). After 5 years of follow-up (median 42 months), freedom from proximal endograft failure was 90.2% (95%CI 85-95) in the PG and 74.9% (95%CI 64-87) in the TG (P=.001). Neck length <7 mm (HR 2.11, 95%CI 1.18-4.56; P<.001) and use of endoanchors as a treatment (HR 3.48, 95%CI 1.61-7.51; P<.001) had a higher risk of proximal endograft failure. In the TG, the use of endoanchors alone (without a proximal aortic extension) (HR 4.00, 95%CI 1.19-8.65; P<.001) and the persistence of a type Ia endoleak at the completion angiography (HR 15.9, 95%CI 10.5-23.0; P<.001) were associated with a higher risk of proximal failure. CONCLUSIONS: In this real-world registry, ESAR was used as an adjunct during primary EVAR in patients with a hostile proximal neck and as a salvage treatment of proximal endograft failure after a previous EVAR, with favorable success rates in both settings. In the preventive setting, outcomes were less favorable when the anatomical proximal neck length was <7 mm. In the treatment of type Ia endoleak, favorable results were obtained in up to 80% of cases, with a higher success rate when endoanchors were combined with a proximal aortic extension, and the persistence of a type Ia endoleak on the completion angiogram was a negative prognostic factor. Future comparative studies are necessary to better clarify the clinical efficacy of ESAR in a prevention and treatment setting.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Efficacy and safety of endovascular aneurysm repair for abdominal aortic aneurysms with hostile proximal neck anatomy: a single-center retrospective cohort studyFrontiers in Cardiovascular Medicine · 2026
- Risk Prediction With a Type Ia-Oriented Proximal Neck Anatomic Severity Grading Score After EVAR.Annals of vascular surgery · 2026
- Outcomes of Partial Endograft Explantation for Noninfectious Failures After Endovascular Aneurysm Repair: A Retrospective Study.Annals of vascular surgery · 2026
- Mechanical Characterization and Ex Vivo Evaluation of a Novel Constraining Device for Safe Endovascular Endograft Explantation.Annals of vascular surgery · 2026
- Physician-Modified Fenestrated Endovascular Aortic Aneurysm Repair After Failed EVAR with Type I Endoleak Offers Favorable Results Compared to Open Conversion.Annals of vascular surgery · 2026
- Geometrical Analysis of Target Vessel Instability After FEVAR With Anaconda Stent-Graft SystemJournal of Endovascular Therapy · 2026