Influence of Cardiac Phase on Valve-to-Coronary Ostia Distance in Valve-in-Valve Transcatheter Aortic Valve Replacement.
Farah Cadour, Felipe Castillo, Mini Pakkal, Felipe Soares Torres, Jaydeep Halankar
Joint Department of Medical Imaging, University Medical Imaging Toronto, Toronto General Hospital, University Health Network, University of Toronto, ON, Canada.
Key points
From the paper's conclusion- VTC measurements differed and were slightly higher in systole than in diastole, which may influence the coronary obstruction risk classification.
Abstract
PURPOSE: No consensus exists on which cardiac phase to use for virtual valve-to-coronary ostium (VTC) distance measurement to predict risk of coronary obstruction in valve-in-valve (ViV). Our study aimed to investigate whether cardiac phase influences VTC measurements in ViV for failing stented surgical heart valves (SHV). METHODS: All adult patients with stented SHV referred for a pre-planning CT for ViV between 2020 and 2025 were retrospectively eligible. For each patient, VTC was measured for two transcatheter heart valve options at both the right and left coronary ostia. VTC measurements were compared between systole, diastole, and phase with best image quality (best phase). Prevalence of changes in coronary obstruction risk (based on 4-mm cutoff) depending on the phase was assessed. Inter- and intraobserver reliability was evaluated. RESULTS: A total of 398 VTC measurements derived from 116 included patients (69 male, 71 ± 13 years) were evaluated. Best phase was more often diastolic. Median VTCs differed between systole and diastole (P < .001), and systole and best phase (P < .001), but not between diastole and best phase (P = .47) (systole 4.8 [3.1-6.5] mm vs diastole 4.6 [3.1-6.2] mm vs best phase 3.9 [2.5-6.2] mm). Measurements were slightly higher in systole than in diastole (mean difference +0.24 mm, [95% CI: 0.16, 0.32]; P < .001), but similar between diastole and best phase (P = .47). Risk of coronary obstruction changed between systole and diastole in 41 coronary arteries (41/398, 10%). Intra- and interobserver reliability was good-to-excellent. CONCLUSIONS: VTC measurements differed and were slightly higher in systole than in diastole, which may influence the coronary obstruction risk classification.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Quantifying the impact of coronary-related experimental simplifications on flow features in native and neo-sinus following transcatheter aortic valve implantation.Biomechanics and modeling in mechanobiology · 2026
- Phase-dependent agreement between noninvasive and invasive blood pressure monitoring during transcatheter aortic valve replacement: a prospective observational study.Canadian journal of anaesthesia = Journal canadien d'anesthesie · 2026
- Real-Time Hemodynamic Guidance During Valve-in-Valve TAVI.JACC. Case reports · 2026
- UNICORN for Valve-in-Valve Transcatheter Aortic Valve Replacement: UNICORN Hong Kong Registry.JACC. Asia · 2026
- In-Vivo Frame Dynamicity of Self-Expanding Transcatheter Aortic Valves.Catheterization and cardiovascular interventions : official journal of the Society for Cardiac Angiography & Interventions · 2026
- Agreement of noninvasive multimodality imaging measurements with balloon sizing in the cardiac catheterization laboratory prior to percutaneous pulmonary valve implantation.The Journal of invasive cardiology · 2026